PLJ-N-BA-e-1333
21
7.2.4 Time setting to start automatic internal adjustment
– „CAL t“
Note
:
This adjustment function is locked [Er 9 lock] in verifiable devices.
Set the time interval after which automatic adjustment is to be started in this menu
item as follows:
Operation:
Display:
Start balance by pressing
0.0000 g
Press
.
P1 CAL
Press
.
P1.1 iCAL
Press
repeatedly until function "
P1.5 CALt
“ appears
P1.5 CALt
Press
, current setting will be flashing
1 h
Use the arrow keys (
) to select a time interval 1
– 12 h
12 h
Confirm setting by pressing
. Weighing balance returns to
menu.
P1.5 CALt
Press
repeatedly until “SAVE“? appears.
SAVE ?
The changes you made will be saved by pressing
.
To cancel changes, press the
key. Afterwards the balance
automatically jumps back to weighing mode.
0.0000 g