11 Troubleshooting
The balance should be switched off for a short time following an interruption in the
programme sequence and disconnected from the mains supply. It is then necessary
to repeat the weighing process from the beginning.
Help:
Malfunction Possible
cause
Weight display is not
illuminated.
•
The balance is not switched on.
•
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
•
Power supply interrupted.
•
Batteries are inserted incorrectly or empty
•
There are no batteries inserted.
The weight display changes
continually
•
Draught/air movement
•
Table/floor vibrations
•
Weighing plate has contact with other objects.
•
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is
obviously wrong
•
The display of the balance is not at zero
•
Adjustment is no longer correct.
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Switch the balance off if other error messages should appear and then switch on
again. If error message persists, inform manufacturer.