7.3.1.1 Calibration (adjustment) only for model KERN ALT 310
Operator
indication
Ö
Call up menu item
“P1 Calibration“
(see chapter 7.1)
Ö
Press
-key
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
Setup
2.02.05
01 Int. cal. - Adjustment with internal weight
Ö
Use the cursor (
►
) to select
“01 Int. cal.“
Setup
P 1 01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
fun c tio n
fun c tio n
fun c tio n
o ff
o n
fun c tio n
31.01.05
1 1 h o u r
Ö
Press the key; the adjustment with internal
weight is carried out.
Calibration
Internal calibration
Ö
After successful adjustment the balance
automatically returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing plate) the display will show an error
message, repeat adjustment.
Weighing
0.0000
g
0%
100%
0
27.01.05
02 Ext. Cal - Adjustment with an external weight
Ö
Use the cursor (
►
) to select “
02 Ext. cal.“
Setup
P 1 01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
fun c tio n
fun c tio n
fun c tio n
o ff
o n
fun c tio n
31.01.05
1 1 h o u r
Ö
Press -key
Ensure that there are no objects on the weighing
plate.
Ö
Press -key; the display shows the measuring
value of the adjusting weight.