ALT-BA-defsi-0410
83
Ö
After successful adjustment the balance
automatically returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing plate) the display will show an error
message, repeat adjustment.
Weighing
0.0000
g
0%
100%
0
27.01.05
02 Ext. Cal - Adjustment with an external weight
Locked for verifiable appliances
03 User cal.
– locked for verifiable appliances
04 Cal. test
– not documented
05 Weight corr.
– locked for verifiable appliances
06 Autom. cal.
– locked for verifiable appliances
07 Auto. Cal.time
– locked for verifiable appliances
08 Print report
– printout of the adjustment data
Ö
Use the cursor (
►
) to select
“08 “print report“
Ö
Press the
–key; the current menu item blinks.
Use the arrow keys (see chapter 7.1) to save your
settings
0 Print
report
deactivated
1 Print
report
activated
Setup
P 1 01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
fun c tio n
fun c tio n
fun c tio n
o ff
o n
fun c tio n
31.01.05
1 1 h o u r
Ö
Confirm with -key
Ö
When
print report
is activated you will receive a printout of your adjustment data
after each adjustment (Ex.):
*** Internal calibration report ***
Calibr. :
internal
Diff..
:
0.0038 g
Signature………………………….