CB/FCB-BA-e-0640
20
7.11 Animal weighing function
The animal weighing function can be applied for busy weighings. This setting finds
out an average weighing value over a defined period of time.
In the menu the time can be set which is necessary to find out the average value.
Proceed as follows:
Switch-on the balance using the
key and wait for the
„0“
display.
Press the
key and keep it pressed until
„UNIT“
appears on the display.
Press the
key eight times, in the display appears
„ANL“
.
Confirm by pressing the
key. Use the
key to choose between the
following settings:
Display
Adjustment
„ANL“ 3
Average value found out after 3 seconds
„ANL“ 5
Average value found out after 5 seconds
„ANL“ 10
Average value found out after 10 seconds
„ANL“ 15
Average value found out after 15 seconds
„ANL“ Off
Animal weighing not active
Use the
key to confirm the selected setting.
Operation:
Switch-on the balance using the
ON
key and wait for the
„0“
display.
Put the weighing good (animal) on the weighing plate and press the
key. In the
display the preselected time is displayed in seconds and then is counted towards
zero. During this time the balance takes up several measuring values. When
reaching
„0“
sounds an acoustic signal and the calculated weighing value is
displayed.
By pressing the
key several times, the balance returns to the normal weighing
mode.
Repeated pressing of the
key activates this function anew.