background image

CB/FCB-BA-e-0640 

12

 

If the batteries are exhausted, „

LO

“ is displayed; press 

 and replace the batteries 

immediately. 
If the balance is not used for a longer time, take out the batteries and store them 
separately. Leaking battery liquid could damage the balance. 
 
If there exists an optional rechargeable battery, it has to be connected in the battery 
compartment via a separate plug-in socket. Now the mains adapter delivered with the 
rechargeable battery must be applied. 
 

6.5  Connection of peripheral devices   

Before connecting or disconnecting of additional devices (printer, PC) to the data 
interface, always disconnect the balance from the power supply. 
With your balance, only use accessories and peripheral devices by KERN, as they 
are ideally tuned to your balance. 

6.6 Initial 

Commissioning 

In order to obtain exact results with the electronic balances, your balance must have 
reached the operating temperature (see warming up time chap. 1). During this 
warming up time the balance must be connected to the power supply (mains, 
accumulator or battery). 
The accuracy of the balance depends on the local acceleration of gravity.  
Strictly observe hints in chapter Adjustment. 

6.7 Adjustment 

As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated - in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). This 
adjustment process must be carried out for the first commissioning, after each 
change of location as well as in case of fluctuating environment temperature. To 
receive accurate measuring values it is also recommended to adjust the balance 
periodically in weighing operation. 

6.8 Adjustment 

The adjustment should be made with the recommended adjustment weight (see 
chap. 1 "Technical data"). Adjustment is also possible with the weights of other 
nominal values (see table 1), but not the optimum for measuring technique.  
 

Procedure when adjusting: 

Observe stable environmental conditions. A warming up time (see chapter 1) is 
required for stabilization. 

Switch on the balance using the 

 key. 

Press 

 key and keep it pressed, after the acoustic signal in the display appears 

for short time „

CAL

“. After that the exact size appears flashing in the display 

(chapter.8.4) of the adjustment weight. 

Now set the adjusting weight in the centre of the weighing plate. 

Содержание CB3K0.1N

Страница 1: ...Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating instruction Compact balances KERN CB FCB Version 4 0 12 2006 GB C...

Страница 2: ...aging__________________________________________________ 9 6 Unpacking Setup and Commissioning ____________________________ 10 6 1 Installation Site Location of Use ______________________________ 10 6...

Страница 3: ...weight ____________________________ 24 8 5 Interface RS232C ___________________________________________ 25 8 5 1 Data transfer mode ________________________________________ 25 8 5 2 Baud rate _______...

Страница 4: ...g 1 g 2 g Warm up time 2 hours 30 minutes 10 minutes Reference quantities at piece counting 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class...

Страница 5: ...up time 30 minutes 30 minutes 10 minutes Reference quantities at piece counting 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details for...

Страница 6: ...arm up time 30 minutes 30 minutes 10 minutes Reference quantities at piece counting 5 10 25 50 Weighing Units Details Weighing units chapter 7 9 Recommended adjustment weight not added class Details f...

Страница 7: ...ive about electromagnetic compatibility in the version 89 336 EWG Applied harmonized standards in particular EN 55022 1998 EN 61000 3 2 1995 A1 1998 A2 1998 A14 2000 EN 61000 3 3 1995 EN 55024 1998 An...

Страница 8: ...This may damage the measuring system Impacts and overloading exceeding the stated maximum load max of the balance minus a possibly existing tare load must be strictly avoided Balance may be damage by...

Страница 9: ...ety Precautions 4 1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning even if you are already familiar with KERN balances 4...

Страница 10: ...se the device to extreme dampness for longer periods of time Non permitted condensation condensation of air humidity on the appliance may occur if a cold appliance is taken to a considerably warmer en...

Страница 11: ...4 Operation using a rechargeable battery optional Lift off the battery cover on the lower side of the balance Connect 9 V compound battery Reinsert the battery cover For battery operation the balance...

Страница 12: ...observe hints in chapter Adjustment 6 7 Adjustment As the acceleration value due to gravity is not the same at every location on earth each balance must be coordinated in compliance with the underlyin...

Страница 13: ...of the adjustment weight An error during adjustment or the use of an incorrect adjusting weight will result in an error message CAL E Repeat adjustment Keep the adjustment close to the balance Daily c...

Страница 14: ...now internally saved If after finishing the weighing process the key is pressed again 0 appears anew in the display The taring process can be repeated any number of times e g when adding several compo...

Страница 15: ...key and wait for the 0 display Press key shortly Appears the reference unit number 5 By pressing the key several times other reference quantities 10 25 and 50 can be called up Place as many pieces to...

Страница 16: ...e key Weigh component p when pressing the key appears the net total i e the sum weight of components n and o and p If necessary also fill the formula up to the desired final value Back to weighing mod...

Страница 17: ...962 Troy ounce ozt 0 032150747 Tael Hongkong tlh 0 02671725 Tael Taiwan tlt 0 0266666 Pennyweight dwt 0 643014931 Momme mom 0 2667 Tola tol 0 0857333381 Freely selectable factor FFA xx xx In order to...

Страница 18: ...0 5N CB3K1N FCB6K1 FCB12K1 FCB15K5 FCB24K1 FCB24K2 FCB24K10 Gram X X X X X X X X X kg X X X X X X X Pound X X X X X X X X X Ounce X X X X X X X X X Troy ounce X X X X X X X X X Tael Hongkong X X X X X...

Страница 19: ...Press the key seven times in the display appears bl Confirm by pressing the key Use the key to choose between the following three settings Display Adjustment Function bl on Background illumination on...

Страница 20: ...age value found out after 3 seconds ANL 5 Average value found out after 5 seconds ANL 10 Average value found out after 10 seconds ANL 15 Average value found out after 15 seconds ANL Off Animal weighin...

Страница 21: ...t for the 0 display Press the key and keep it pressed until UNIT appears on the display Press the key nine times in the display appears rSt Confirm by pressing the key Use the key to choose between th...

Страница 22: ...c until UNIT appears By pressing the key the different menu items are called up Use the key to select a menu item Within this menu item use the key to make your choice If the key is repeatedly actuate...

Страница 23: ...ally In the event that small quantities are removed or added to the material to be weighed incorrect weighing results can be displayed due to the stability compensation in the balance Example Slowly d...

Страница 24: ...er to achieve high quality weighing results in the sense of the measuring technology it is recommended to select the nominal value as high as possible The non delivered adjustment weights can be purch...

Страница 25: ...de can be converted using the key Pr PC AU PC AU Pr re Cr BA Pr details chapter 9 4 AU Pr 5 The changed setting is taken over by pressing the key 6 The balance returns to the weighing mode 0 0 g 8 5 2...

Страница 26: ...parameters using the key to scroll backward use the key After actuating the key the current state is displayed on off Use the or the key to change the state After that the current state is saved using...

Страница 27: ...nation of the data transfer 9 3 1 Pr PC Press the PRINT key at stable weight the format is transferred from LAPR a Format for stable values for weight quantity percentage 1 2 3 4 5 6 7 8 9 10 11 12 13...

Страница 28: ...ASCII code After the balance having received the s w t commands it will send the following data Take into account that the following remote control commands must be sent without a subsequent CR LF s F...

Страница 29: ...ded Take into account that the output format of the balance is fixedly defined and cannot be changed The printer format is stored in the printer i e in case of a failure the printer cannot be changed...

Страница 30: ...and wipe with a dry soft cloth Loose residue sample powder can be removed carefully with a brush or manual vacuum cleaner Spilled weighing goods must be removed immediately 10 2 Service maintenance Th...

Страница 31: ...y No batteries inserted The displayed weight is permanently changing Draught air movement Table floor vibrations The weighing plate is in contact with foreign matter Electromagnetic fields static char...

Отзывы: