CB/FCB-BA-e-0640
15
7.5 Plus/minus
weighings
For example unit weight control, fabrication control etc.
Switch-on the balance using the
key and wait for the
„0“
display.
Put the nominal weight on the weighing plate and tare to „
0
“ using the
key. Remove the nominal weight.
Put the test objects subsequently on the weighing plate, the respective deviation from
the nominal weight is displayed with the respective sign to „+“ and „-“.
According to the same procedure also packages with the same weight can be
produced, referring to a nominal weight.
Back to weighing mode by pressing the
key.
7.6 Parts counting
Switch-on the balance using the
key and wait for the
„0“
display.
Press
key shortly. Appears the reference unit number
5
.
By pressing the
key several times other reference quantities
10, 25
and
50
can
be called up.
Place as many pieces to count on the weighing plate as the set reference quantity
requires.
Confirm with
key.
The balance is now in parts counting mode counting all units on the weighing plate.
By pressing the
key the balance returns to the weighing mode and displays the
weight of the counted units.
Important: The larger the reference quantity, the more accurate the parts
counting.
Smallest counted weight see table „Technical data“, if this weight is less, in the
display appears „
Er 1
“. Use
key to return to the weighing mode.
The tare vessels can also be used for piece counting. Before starting percent
weighing use the
key to tare out the container.