13 Description of individual menu items
13.1 Automatic cut-off of display backlight
(Auto backlight OFF)
Menu item:
[
B. AB
]
see chpt. 10.1.2
To save battery when function is enabled backlight of display will automatically
switch off after 3 minutes during which there was no change of load or activity.
13.2 How to change weighing units
Menu item:
[
Cq W.A
]
and
[
Cs W.B
]
see chpt. 10.1.2.
This setting can be used to set different display units (A or B) for a weighing value.
Use the
F
-key to choose between units A and B.
•
“Units A“ are available in all operating modes.
•
“Units B“ are only available in weighing mode
[
q SET 1
]
.
13.3 How to change readability (models AES only)
Menu item:
[
]
and
[
]
see chpt. 10.1.2.
This function is used to change readability for weighing unit A or B. The last decimal
place will be rounded or removed from the display.
Example for weighing unit [ g ] :
Function 1
Readability
Reaction to ambient con-
ditions
1 0.0001
g
sensitive
2 0.0002
g
3 0.0005
g
4 0.001
g
Menu settings
5 0.002
g insensitive