AES/AEJ-BA-e-0912
24
9.4 Adjustment with external weight (only AES)
During adjustment tests the balance automatically compares the saved value of the
adjustment weight with the actual value. This is only a check, i.e. no values are
changed.
Condition
: For menu setting [
x CA. 4
] see chpt. 10.1.2
Ö
Press the
CAL
key
T. EHT.
CAL
1
PWSH C
CAL
Ö
Operate the
CAL-
key
T. EHT.
CAL
1
Qn. O
CAL
Ö
When display
[
on F.S
]
appears, put the adjusting
weight carefully in the centre of the weighing pan.
CAL
Qn
F.5.
Ö
Adjustment test is starting
CAL
Qn
F.5.
Ö
The difference between the saved value and the
measured value is displayed.
Ö
Adjustment protocols will be issued if you connect
an optional printer and by pressing the
-
key, see chpt. 15 “printer mode”. Whilst printing is
in progress [
] will be displayed.
D1FF
CAL
z0007
CAL
g
Ö
Press any key; adjustment process will stop. Re-
move adjusting weight, balance will return into
weighing mode.
EnD
CAL