AES/AEJ-BA-e-0912
23
9.3 Adjustment with external weight (only AES)
Carry out adjustment with the help of the recommended adjustment weight (see
chapter 1 “Technical Specifications”).
Info about adjustment weights can be found on the Internet at: http://www.kern-
sohn.com
Condition
: Menu setting [
x CA. 3
], see chpt. 10.1.2
Ö
Press the
CAL
key
EHT.
CAL
CAL
1
K
PWSH C
CAL
Ö
Press the
CAL
-key again; zero point will be
saved.
EHT.
CAL
CAL
1
K
Qn. O
CAL
Ö
When display
[
on F.S
]
appears, put the adjusting
weight carefully in the centre of the weighing pan.
CAL
Qn
F.5.
Ö
The adjustment process is started
CAL
Qn
F.5.
Ö
The process of adjustment is completed.
Adjustment protocols will be issued automatically
if you connect an optional printer, see chpt. 15
“printer mode”. Whilst printing is in progress
[busy
] will be displayed.
BWSX
CAL
EnD
CAL
Ö
The balance returns automatically into weighing
mode.
Take away adjustment weight. An adjusting error
or incorrect adjusting weight will be indicated by
the error message
[- Err
]
. Repeat adjustment
procedure.