background image

13

The BATRT-M is provided with extensive self checking which assists the user in the location of setup entry errors 

and in reporting malfunctions or unusual operating conditions. When an error occurs, the display will flash.  Press 

any key to see the error message corresponding to the error that has occurred.  Press any key again to acknowl-

edge the error. (If the error can be eliminated by a change of setup values, the unit will automatically advance to 

the MENU so that the appropriate setup changes can be made).  

 Table - 2 illustrates the warning message, problem, and recommended corrective actions.

Diagnostic Error Messages Table-2 

WARNING MESSAGE 

CAUSE 

CORRECTIVE ACTION

RATE ER

 

Rate Low set  

Set Rate Hi greater

 

higher than Rate Hi     than Rate Lo 

 

FAC ERR

 

Factor = 0 

Set in correct Factor

“BAT” Descriptor 

Low Battery 

Replace battery(ies)

E total

 

Total rollover 

None required

E RATE

 

Rate exceeds 99999 

Use lower rate dec point

E pulse

 

Pulse out Overflow 

Use different pulse scaler or 

 

 

totalizer decimal point

E FLASH

 

Save to flash  

Write down displayed total and 

 

memory failed 

setup values if you are changing 

 

 

the battery.  If total wasn't saved, 

 

 

it will display an arbitrary total 

 

 

when new battery is installed.  In 

 

 

this case, reset the total to 0 and 

 

 

check the setup information.

  

ANALOG OUTPUT CALIBRATION

CAUTION: Performing the analog output calibration will erase all programming values.  Please record all 

programmed values before beginning the analog output calibration
If the unit is equipped with the analog output option, 
the 4-20 mA has been accurately set to 4.000 and 

20.000 mA by the factory. No calibration should be required.
The 4-20 mA output may be verified  periodically by installing a digital milliamp meter(DMM) in series with the 

analog output and simulating a full scale or over range flow rate.
If the output is out of calibration, perform the following:

Remove power and/or disconnect battery(ies). Hold down the scroll (

) key and replace the battery.  This will 

initialize the unit and advance to the analog output calibration mode. Connect a DMM set to read current in series 

with a power supply (8.5 to 30 volts DC) to TB12 (+) and TB11 (–).  The output should read 4.000 ma (± 0.005).  

If it does not, Adjust the numbers on the display up or down until the output reads 4.000 ma (± 0.005).  Press the 

"E" key. The DMM should now read 20.000 ma (± 0.005).  If it does not, Adjust the numbers on the display up or 

down until the output reads 20.000 ma (± 0.005).  Press "E"  and the unit will return to the "RUN" mode.

ERROR MESSAGES

Содержание BAT R/T Millennium-S1

Страница 1: ...Ratemeter Totalizer with RS232 Serial Port Installation and Operating Instructions 99982 11 10 15 KESSLER ELLIS PRODUCTS 10 Industrial Way East Eatontown NJ 07724 800 631 2165 732 935 1320 Fax 732 935...

Страница 2: ......

Страница 3: ...Installation and Replacement 3 Dimensions 4 Theory of Operation 5 Wiring 5 Typical Applications 5 Definitions 6 Programming Flowchart 9 Operation 12 Error Messages 13 Analog Output Calibration 13 War...

Страница 4: ......

Страница 5: ...MOSPHERE DE C DE GROUPE AVERTIR BOUCHER LES ENTREES INUTILISEES DE CONDUIT AVEC UN BOUCHON DE CONDUIT CERTIFIE ENVIRONMENTAL OPERATING TEMPERATURE 4 F 20 C to 158 F 70 C Extended Temp 22 F 30 C to 158...

Страница 6: ...battery life by 20 IDATA STORAGE Setup Information Stored in flash memory Totalizer Stored in battery backed RAM but can be saved to flash memory by operator for recall after battery change out COMMU...

Страница 7: ...iptor BAT illuminates when the battery voltage falls below this predetermined value The low battery detector operates correctly with all power options The battery or batteries should be replaced withi...

Страница 8: ...RT M 0 BATRT M 1 E M 3 0120 DIAMETER 900 22 8 992 25 2 Mounting holes 125 Drill 2 places 438 Nylon Spacer Supplied 1 8 46 2 875 73 10 2 5 438 11 1 900 22 8 992 25 2 2 40 61 2 20 56 0 95 24 0 30 max 7...

Страница 9: ...t frequency is used to access the table A linear interpolation of adjacent point pairs is used to arrive at the K Factor at that input frequency The flow rate and total are then computed based upon th...

Страница 10: ...t is linear over its entire operating flow range Press the key to step to the desired choice Press the E key to enter the displayed factor type No yes set Pnts set 20 point This prompt allows the user...

Страница 11: ...ay r deCloC ratemeter decimal location Sets the decimal location for the ratemeter information The ratemeter deci mal is not a dummy decimal and will scale the rate display accordingly i e if the r de...

Страница 12: ...g display will result Pressing the M key will result in a display of an error mes sage E PULSE Press the E key to return to the run mode P uuidth pulse width Sets the pulse width of the pulse output S...

Страница 13: ...ep the factor decimal to the desired location Press the E key to enter the displayed decimal location Press the M key to skip and keep the existing location NOTE For best performance and resolution ch...

Страница 14: ...ress the E key to enter the displayed value Press the M key to skip and keep the existing value Press the key to increment each individual digit of the out high setting 20mA value This will only displ...

Страница 15: ...setting Press the E key to enter the displayed lock setting Press the M key to skip and keep the existing setting Press the key to increment each individual digit of the lock code Press the key to ad...

Страница 16: ...panel switch sequence or by a contact closure on the remote reset terminal to circuit common To reset the unit from the front panel the following key sequence is required Press M CLr tot will be displ...

Страница 17: ...display an arbitrary total when new battery is installed In this case reset the total to 0 and check the setup information ANALOG OUTPUT CALIBRATION CAUTION Performing the analog output calibration wi...

Страница 18: ...y 4 20 Point Linearization 10 point with S2 option D Rate per Day Hour or Minute ET Extended Temp 22 F to 158 F 30 C to 70 C CE CE Approved Version LVD EMC only CSA CSA Approved Version explosion proo...

Страница 19: ...15 Revision History 03 09 05 Original release 11 10 15 Added Specifications and Dimensions for 3NE and 3SS mounting options...

Отзывы: