10
PROGRAMMING FLOWCHART
(continued)
Press the key to step the totalizer decimal to the desired location.
Press the
E
key to enter the displayed decimal location.
Press the
M
key to skip and keep the existing location
Press the key to step to the desired totalizer descriptor.
Press the
E
key to enter the displayed descriptor.
Press the
M
key to skip and keep the existing descriptor.
NOTE:
When option "D" (rate per day) is ordered; selections are:
GAL, BBL, MCF, M3, "blank"
Press the key to step to the desired scale setting.
Press the
E
key to enter the displayed scale setting.
Press the
M
key to skip and keep the existing setting.
NOTE:
When option "D" (rate per day) is ordered; selections are:
min, HRS, days
Press the key to step the ratemeter decimal to the desired location.
Press the
E
key to enter the displayed decimal location.
Press the
M
key to skip and keep the existing location
Press the key to increment each individual digit of the delay.
Press the key to advance to the next digit.
Press the
E
key to enter the displayed value.
Press the
M
key to skip and keep the existing value.
Press the key to increment each individual digit of the out low setting (4mA
value).
This will only display on units with Analog Output.
Press the key to advance to the next digit.
Press the
E
key to enter the displayed value.
Press the
M
key to skip and keep the existing value.
Press the key to increment each individual digit of the out high setting
(20mA value).
This will only display on units with Analog Output.
Press the key to advance to the next digit.
Press the
E
key to enter the displayed value.
Press the
M
key to skip and keep the existing value.
Press the key to step to the desired pulse scale divider for the pulse output.
Press the
E
key to enter the displayed pulse scale divider.
Press the
M
key to skip and keep the existing pulse scale value.
Press the key to step to the desired pulse width for the pulse output.
Press the
E
key to enter the displayed pulse width.
Press the
M
key to skip and keep the existing pulse width.
This will not display if Pulse Out is turned OFF.
Press the key to increment each individual digit of the "nor" factor.
Press the
E
key to enter the displayed value.
Press the
M
key to skip and keep the existing value.
TDEC
12345678
ToT DESC
SEC
r SCALE
SEC
NNIN
HRS
(
Days
)
GAL
LIT
FT3
M3
"blank"
(BBL, MCF)
1234
R DECLoC
E
M
E
M
E
M
E
M
0.1
delay
0000
oUT Lo
99999
oUT HI
D 1
Pul scale
D 1
D 10
D100
oFF
D 1
D 10
D100
oFF
0.1
to
99.9
E
M
E
M
E
M
E
M
.5
p uuidth
E
M
.5
.25
.125
.0625
1
nor
E
M
Continued From Previous Page
Continue On
Next Page
TOTAL DECIMAL
TOTAL
DESCRIPTOR
SCALE
(ratemeter)
RATE
DECIMAL
LOCATION
DELAY
OUT LOW
(4mA)
OUT HIGH
(20mA)
PULSE
SCALER
(divider)
PULSE
WIDTH
(seconds)
NORMALIZATION
(dampening)