25
EN
Operating instructions CellaTemp PK / PKF / PKL xx
The hold time [
PhLd
] can be set from 0�1 to 999�9 sec� The maximum tempera-
ture sampled during the defined hold time will be saved� It makes sense to choose
a hold time which is approximately 1�5 times as long as the cycle of the moving
targets� This avoids temperature drops� Any changes are recognised at once�
1
2
3
4
5
OUT2
[mA]
t
1: Measuring object in front of pyrometer
2: Hold time
3: second internal hold time
4: Measuring readings with peak hold function
5: Measuring readings without peak hold function
13.5 Reset all parameters to factory settings
►
[
rES.
] select in menu advanced functions
►
Press [ENTER]
> The display shows [STOP]
►
Hold [ENTER] for 2 seconds
> Display flashes for 2 seconds
► Press [▲]
> The display shows [
EHEc.
]
►
Press [ENTER]
> The display shows the current temperature
Содержание CellaTemp PKF Series