49
EN
Operating instructions CellaCast PX 8x
17 Service functions
17.1 Reset all parameters to factory settings
17.2 Simulate a desired temperature
►
[
rES.
] Select extended functions [
EF
] in the menu
►
Press [MODE]
> RES is shown in the display
► Press the ˅ key for 2 s
> RES flashed for 2 s
► Release the key ˅ key and press it again
> – – – – is shown in the display
►
Press [MODE] button
> The current temperature is shown in the display
►
[
t.SiM
] Select extended functions [
EF
]
►
Press [MODE]
> The display shows the previously set temperature
► Press the ˄ key for 2 s
> Temperature flashes 3 times
► Use the keys ˄ ˅ to set the desired temperature, release the key
> The display alternately shows
t.SiM
and the temperature value
►
Press [MODE] button
> The display shows
t.SiM
and the simulation is completed
Exit operating parameter layer
►
Wait 30 seconds
or
► Press [˄ or ˅] to change to the parameter End. Then press [MODE] to change
to the functional menu�
► In the functional menu press [˄ or ˅] to change to the parameter
End
,
then press [MODE]�