Operating instructions CellaCast PX 8x
10 Adaptation of temperature level
By changing the emissivity ratio, the difference between the measured temperature
level and true temperature can be compensated�
For this purpose, a comparison measurement with an immersion lance is required�
Adjust emissivity ratio in such a way that the measured values coincide� An increase
in the emissivity ratio results in a low temperature value�
The graph shows an approximation of the influence of the emissivity ratio setting on
the change in temperature value�
1 Ratio correction
2 Temperature difference [°C]
Experience has shown that the following settings apply as a first approximation
- grey pig iron 101
- spheroidal graphite iron 104
However, these values may differ in practice� After changing the emissivity ratio,
carry out a new control measurement�
The emissivity ratio is set as follows: