Operating instructions CellaCast PX 8x
choose a hold time which is approximate 1�5 time as long as cycle of the moving
targets� This avoids temperature drops� Any changes are recognised at once�
Two-colour mode [
q
] → [
q.Phd
] → [
t iME
]
One-colour mode [
L1
] →
[
'.PhD
]
→ [
t iME
] or
[
L2
] →
[
".PhD
]
→ [
t iME
]
12.13 ATD function
This function is used to automatically measure the temperature of a discontinuous-
ly running process� For example, it can be used to determine the temperature of
bolts passing the pyrometer acyclically and which have a variable length�
The start of a measuring cycle is determined automatically and is dependent on
the following variables:
t
Ao
[mA]
1
2
3
4
5
1 Measuring object in front of the pyrometer
2 Hold time
3 Second internal hold time
4 Measuring reading with peak hold function
5 Measuring reading without peak hold function