Models 2510 and 2510-AT User’s Manual
Basic Operation
3-11
Step 5: Turn output on.
Turn the output on by pressing the ON/OFF OUTPUT key. The OUTPUT indicator will
turn on to indicate the output is on.
Step 6: Observe the display.
During operation, you can observe the front panel display for operating and measured
parameters. For example, for the temperature function, the display will show the measured
temperature and the temperature setpoint, and you can use the DISPLAY TOGGLE key to
cycle through the TEC voltage, current, power, and resistance values, as well as the sensor
resistance.
Step 7: Turn output off.
When finished, turn the output off by pressing the ON/OFF OUTPUT key. The OUTPUT
indicator light will turn off to show the output is off.
Basic remote control-measure procedure
Use the following procedure to perform the basic control-measure operations via remote.
See
for command details and
for an example program.
NOTE
The following procedure assumes that the Model 2510 is already connected to
the thermoelectric cooler and temperature sensor as explained in
Step 1: Restore defaults.
Send the following command to restore defaults:
*RST
See
Step 2: Select function.
Send the :SOUR:FUNC command to select the desired control function. For example, to
select the temperature function, send the following command:
:SOUR:FUNC TEMP