Models 2510 and 2510-AT User’s Manual
SCPI Command Reference
10-43
NOTE
Bit 12 of the Measurement Event Register will be set when the setpoint is within
the percent tolerance for the programmed number of readings (see below). See
for status register details.
:COUNt <n>
:SOURce[1]:STOLerance:COUNt <n>
Set number of readings for tolerance
Parameters
<n> =
1 to 100
Number of readings
MINimum
1
MAXimum
100
DEFault
5
Query
:COUNt?
Query number of readings.
Description
This command is used to program how many readings should be within
the tolerance window programmed with the :PERCent command for the
setpoint to be considered to be within tolerance.
:POINts?
:SOURce[1]:STOLerance:POINts?
Query number of readings within window
Description
This command queries the number of readings that have currently been
within the programmed tolerance window.
PID autotune (Model 2510-AT only)
Select minimum settling time criteria
:MSETtle
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MSETtle:GAIN?
Query gain constant
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MSETtle:DERivative? Query derivative constant
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MSETtle:INTegral?
Query integral constant
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MSETtle:TRANsfer
Set constants as defaults
Description
These commands are used to control PID autotuning for minimum set-
tling time criteria. (See
.) The GAIN, DER, and INT queries
request the autotune loop constant values while the TRAN command
transfers constants as default values.