10-44
SCPI Command Reference
Models 2510 and 2510-AT User’s Manual
Select minimum overshoot criteria
:MOVershoot
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MOVershoot:GAIN?
Query gain constant
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MOVershoot:DERivative? Query derivative constant
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MOVershoot:INTegral?
Query integral constant
:SOURce[1]:TEMPerature:ATUNe:LCONstants:MOVershoot:TRANsfer
Set constants as defaults
Description
These commands are used to control PID autotuning for minimum over-
shoot criteria. (See
.) The GAIN, DER, and INT queries
request the autotune loop constant values while the TRAN command
transfers constants as default values.
Query tau and lag values
:SOURce[1]:TEMPerature:ATUNe:TAU?
Query tau value
:SOURce[1]:TEMPerature:ATUNe:LAG?
Query lag constant
Description
These queries are used to query the autotune tau and lag values.
Set temperature start and stop values
:SOURce[1]:TEMPerature:ATUNe:STARt <n>
Set start temperature
:SOURce[1]:TEMPerature:ATUNe:STOP <n>
Set stop temperature
Description
These commands are used to control the start and stop temperatures for
PID autotuning (See
Parameters
<n> =
-50 to +225˚C
223.15 to 498.15K
-58 to 437˚F
Query
:STARt?
Query start temperature
:STOP?
Query stop temperature
Description
These commands are used to set the start and stop temperatures for the
autotune procedure. The Model 2510-AT will use these values to deter-
mine the correct loop constants.
Initiate autotune
:SOURce[1]:TEMPerature:ATUNe:INITiate
Initiate autotune
Description
This command is used to initiate the autotune process. You must first set
the start and stop temperatures before starting the process as described
in