Models 2510 and 2510-AT User’s Manual
SCPI Command Reference
10-35
Current function
[:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude] <n> Program current amplitude
Parameters
<n> =
-5 to +5A
Set current amplitude
MINimum
-5
MAXimum
+5
DEFault
0.0
Query
[:AMPLitude]?
Query current amplitude
Description
This command programs the current amplitude (setpoint) for the current
source function. The current amplitude can be set over a range of ±5A
but cannot exceed the current protection limit.
:LCONstants
:SOURce[1]:CURRent:LCONstants[:GAIN] <n>
Program proportional constant
:SOURce[1]:CURRent:LCONstants:DERvative <n>
Program derivative constant
:SOURce[1]:CURRent:LCONstants:INTegral <n>
Program integral constant
Parameters
:[GAIN] <n> =
0 to 1e5
Set proportional constant
value
MINimum
0
MAXIMUM
1e5
DEFault
10.0
:DERivative <n>= 0 to 1e5
Set derivative constant value
MINimum
0
MAXimum
1e5
DEFault
0.0
:INTegral <n> =
0 to 1e5
Set integral constant value
MINimum
0
MAXimum
1e5
DEFault
20.0
Query
[:GAIN]?
Query proportional constant
:DERivative?
Query derivative constant
:INTegral?
Query integral constant
Description
These commands program the current function PID loop constants,
which must be set to appropriate values for suitable control characteris-
tics. See
appropriate values.