C-12
DriverLINX Configuration Notes
To determine the DriverLINX Logical Channel number for an EXP-1800
Multiplexer channel, use the following formula or refer to the following
table. Note that DriverLINX uses a 0-based numbering scheme for all
analog input channels and “a/d chan” refers to the native physical channel
on a DAS-1800 where you attached the mux.
logical chan = (a/d chan + 1)
×
16 + mux chan
For example, the Logical Channel address for mux channel 4 attached to
analog channel 3 is
(3 + 1)
×
16 + 4 = 68.
To specify multiplexer input channels 1, 2, and 3 for an EXP-1800
connected to native input channel 0, add 16, 32, and 48 to the
channel/gain list.
Table C-3. Table of Logical Channel Numbers for Eight External EXP-1800
Multiplexers
Native
Chan
Mux #
0
MUX
#1
1
MUX
#2
2
MUX
#3
3
MUX
#4
4
MUX
#5
5
MUX
#6
6
MUX
#7
7
MUX
#8
etc.
MUX
Input
Chan #
0
16
32
48
64
80
96
112
128
1
17
33
49
65
81
97
113
129
2
18
34
50
66
82
98
114
130
3
19
35
51
67
83
99
115
131
4
20
36
52
68
84
100
116
132
5
21
37
53
69
85
101
117
133
6
22
38
54
70
86
102
118
134
7
23
39
55
71
87
103
119
135
8
24
40
56
72
88
104
120
136