4-18
Cabling and Wiring
Connecting and Synchronizing Multiple Boards
You can synchronize up to three DAS-1800 Series boards using trigger
and gate signals from the main I/O connector. Each board can run at the
same conversion rate as the other boards in the system or at a different
conversion rate from other boards in the system.
The onboard pacer clock is designed to be tightly coupled with trigger
and gate operations. After each board receives the trigger or gate,
conversions begin within a defined period of time. If each board is
programmed for a different conversion rate, the first conversion on each
board occurs after this time period and subsequent conversions occur at
the programmed rate.
Figure 4-15 shows two connection schemes for synchronizing multiple
boards. Both schemes use the onboard internal pacer clock to time
acquisitions.
Figure 4-15. Two Connection Schemes for Synchronizing Multiple Boards
In scheme 1, you connect the trigger/gate inputs of the three boards
together and supply the trigger or gate input. A/D conversions on each
board start 400 ±100ns from the active edge of the trigger input. All
conversions start within 100 ±100ns of each other from board to board.
When using scheme 1, you can use the onboard internal pacer clock or an
external pacer clock.
Trigger or
Gate
(optional)
TGIN
TGIN
TGOUT
TGIN
TGIN
TGIN
TGIN
Trigger or
Gate
Board 0
Rate a
Board 0
Rate a
Board 1
Rate b
Board 1
Rate b
Board 2
Rate c
Board 2
Rate c