Remote Model 7710 calibration
1. Connect the Model 7710 to the 7797 Calibration/Extender Board (see “
2. With the power off, install the Model 7710/7797 combination in Slot 1, and select the rear inputs with the
INPUTS switch. Allow three minutes for thermal equilibrium.
3. Accurately measure and record the cold temperature of the Model 7710 card surface at the center of the
card.
4. Press in and hold the Model 27xx OPEN key while turning on the power.
5. Enable calibration by sending the :CODE command. For example, the default command is:
CAL:PROT:CODE 'KI0027xx'
6.Initiate calibration by sending the following command:
CAL:PROT:CARD1:INIT
7. Calibrate the Model 7710 with the following command:
CAL:PROT:CARD1:STEP0 <temp>
Here <temp> is the cold calibration temperature (˚C) measured in Step 3.
8. Send the following commands to save calibration and lock out calibration:
CAL:PROT:CARD1:SAVE
CAL:PROT:CARD1:LOCK