Calibration

The following procedures calibrate the temperature sensors on the Model 7710 plug-in cards.

WARNING

The information in this section is intended only for qualified service person-
nel. Do not attempt these procedures unless you are qualified to do so.

Recommended test equipment

In order to calibrate the Model 7710, you will need equipment summarized in 

Table 3

.

Table 3

Recommended equipment for Model 7710 calibration

Extender board connections

The Model 7710 being calibrated should be connected to the 7797 Calibration/Extender Board, 
and the extender board should then be installed in scanner Slot #1. Note that the module being 
calibrated will be external to the Model 27xx to avoid card heating during calibration.

Model 7710 calibration

NOTE

Before calibrating the Model 7710, make sure that power has been removed from 
the card for at least two hours to allow card circuitry to cool down. After turning on 
the power during the calibration procedure, complete the procedure as quickly as 
possible to minimize card heating that could affect calibration accuracy. Allow the 
Model 27xx to warm up for at least two hours.

Front panel Model 7710 calibration

1. Connect the Model 7710 to the Model 7797 Calibration/Extender Board (see “

Extender board connec-

tions

” above).

2. With the power off, install the Model 7710/7797 combination in Slot 1 and select the rear inputs with the 

INPUTS switch. Allow three minutes for thermal equilibrium.

3. Accurately measure and record the cold temperature of the Model 7710 card surface at the center of the 

card with a digital thermometer.

4. Press in and hold the Model 27xx OPEN key while turning on the power.

5. Press SHIFT then TEST, and then display TEST:CALIB with the up or down range key. Press ENTER, 

select RUN, and then enter the appropriate calibration code (default: 0027xx).

6. With NEW CODE? displayed, use the up or down range key to display N, and then press ENTER.

7. Using the up or down range key, select CARD at the CAL:RUN prompt, and then press ENTER.

8. Set the display value to the cold calibration temperature (°C) you measured in Step 3, then press ENTER 

to complete Model 7710 calibration.

Digital Thermometer:

18 to 23˚C, ±0.1˚C

Keithley 7797 Calibration/Extender Board

Содержание 7710

Страница 1: ...Model 7710Multiplexer Card User s Guide PA 847 Rev B 8 03 A G R E A T E R M E A S U R E O F C O N F I D E N C E...

Страница 2: ......

Страница 3: ...that a shock hazard exists when voltage levels greater than 30V RMS 42 4V peak or 60VDC are present A good safety practice is to expect that hazardous voltage is present in any unknown circuit before...

Страница 4: ...electric shock and fire replacement components in mains circuits including the power transformer test leads and input jacks must be purchased from Keithley Instru ments Standard fuses with applicable...

Страница 5: ...rmistor thermocouple Built in cold junction reference for TC temperature Plug in screw terminal connections WARNING Before operating the Model 27xx with an accessory card verify that the card is prope...

Страница 6: ...m through the instrument backplane To prevent damaging the 7710 module or creating a shock hazard the entire test system and all of its inputs should be de rated to 60V DC 42V rms Card configuration F...

Страница 7: ...Channel 20 HI LO Channels 12 19 HI LO Sense HI LO Input Channel 23 2 Pole Closed 4 Pole Open see Notes To Model 27xx Backplane Channel 21 see Note Backplane isolation Channel 22 see Note Backplane iso...

Страница 8: ...DEL 0 Sets trigger delay to 0 seconds FORM DATA ASCII Formats data as an ASCII string N A for 2701 FORM ELEM READ Returns each reading with units only CALC3 OUTP OFF Disables limit outputs CALC3 LIM1...

Страница 9: ...Dim DataSent As Boolean Dim Data As String A Visual Basic function call when a Form program window is opened Private Sub Form_Load TestRunning False End Sub These two functions exit the program and cl...

Страница 10: ...to 2701 MsgBox Cannot connect to LastAddress 48 IP Connect Error ConnectTo2701 False Exit Function End If Connection to 2701 found ConnectTo2701 True End Function Event to capture data from mainframe...

Страница 11: ...e 2701 over the ethernet connection Sub SendKI Cmd As String Send command to mainframe and wait for data sent event DataSent False Winsock1 SendData Cmd Chr 13 Do DoEvents Loop Until DataSent End Sub...

Страница 12: ...ng Dim i As Long i Len Resp Out_str i 1 While i Len Resp Ch Mid Resp i 1 If Ch Then Ch vbCrLf End If Out_str Out_str Ch i i 1 Wend Text1 Text Out_str End Sub This is the main function for the 7710 tes...

Страница 13: ...off limit test SendKI SYST LSYN OFF Turn off line sync SendKI DISP ENAB OFF Turn off display SendKI SYST AZER STAT OFF Turn off auto zero SendKI CALC3 LIM1 STAT OFF ScanList Turn off limits SendKI TRI...

Страница 14: ...lained as follows 4 wire method The 4 wire Kelvin connection method using the 4 function as shown is Figure 2 is generally preferred for low ohms measurements The 4 wire method cancels the effects of...

Страница 15: ...VSHI VSLO V ITEST VM Input Hi Input Lo 2750 DMM with 7710 card Sense Hi Sense Lo RDUT VM ITEST VSHI VSLO RDUT ITEST ITEST VM Virtually no current flows in the high impedance sense circuit due to the h...

Страница 16: ...0 card crosstalk is specified for an AC signal routed through the module to a 50 load Crosstalk is expressed as a dB magnitude at a specified frequency The specification for crosstalk is provided in t...

Страница 17: ...ply the floating volt age to the card Figure 3 Example of an unsafe measurement practice Card handling precautions The solid state relays used on the 7710 card are static sensitive devices Therefore t...

Страница 18: ...isconnects WARNING The information in this section is intended for qualified service personnel Do not attempt to perform this procedure unless qualified to do so Figure 4 shows how to access the screw...

Страница 19: ...re 4 3 If desired pull the appropriate quick disconnect terminal block off the card 4 Using a small flat blade screwdriver loosen terminal screws and install wires as desired refer to Figure 6 Figure...

Страница 20: ...Table 2 on page 18 and affix it to the module cover 8 Close and lock cover NOTE For maximum system performance all measurement cables should be less than 3 meters Step 1 Place flat head screwdriver un...

Страница 21: ...how typical wiring connections for the following types of measurements Thermocouple connections see Figure 8 2 wire and thermistor connections see Figure 9 4 wire and RTD connections see Figure 10 Vol...

Страница 22: ...rmocouple connections Figure 9 2 wire and thermistor connections Channel 1 HI LO Channel 20 HI LO Channels 2 19 Thermocouple Channel 1 HI LO Channel 20 HI LO Channels 2 19 Resistor or Thermistor Topic...

Страница 23: ...them to the cover of the Model 7710 Use this to record connection information and channel descriptions as needed Channel 1 HI LO Channel 10 HI LO Channels 2 9 Channel 11 HI LO Channel 20 HI LO Channel...

Страница 24: ...erminal and chassis or any two terminals when the 7710 is installed Table 2 Connection log Model 7710 Channel Color Description Card Source H L Card Sense H L CH1 H L CH2 H L CH3 H L CH4 H L CH5 H L C...

Страница 25: ...sible to minimize card heating that could affect calibration accuracy Allow the Model 27xx to warm up for at least two hours Front panel Model 7710 calibration 1 Connect the Model 7710 to the Model 77...

Страница 26: ...7710 card surface at the center of the card 4 Press in and hold the Model 27xx OPEN key while turning on the power 5 Enable calibration by sending the CODE command For example the default command is C...

Страница 27: ...61326 1 CONTACT LIFE 1010 operations at cold switching or max signal level guaranteed by design CONTACT RESISTANCE 10 per channel or 5 per conductor Refer to the Model 7710 User s Guide for measuremen...

Страница 28: ...60 11 77 26 GERMANY Landsberger Strasse 65 82110 Germering 089 84 93 07 40 Fax 089 84 93 07 34 GREAT BRITAIN Unit 2 Commerce Park Brunel Road Theale Berkshire RG7 4AB 0118 929 75 00 Fax 0118 929 75 1...

Отзывы: