42
Model 7706 module
(:CALibration:PROTected:CARD1:STEP0)
Purpose
To perform Model 7706 temperature calibration.
Format
:CAL:PROT:CARD1:STEP0 <TEMP>
Parameter
<temp> = Cold calibration temperature (°C)
Description
:STEP0 performs temperature sensor calibration of the Model 7706. The card must be
allowed to cool down to ambient temperature before calibration, and the cold temperature
of the card must be measured and sent as the <temp> parameter during calibration.
NOTE
Before calibrating the temperature on the Model 7706, make sure that power has been removed from the card
for at least two hours to allow card circuitry to cool down. After turning on the power during the calibration
procedure, complete the procedure as quickly as possible to minimize card heating that could affect calibra-
tion accuracy.
Example
:CAL:PROT:CARD1:STEP0 23
Perform 7706 calibration.
Remote error reporting
Methods to detect and determine the nature of calibration errors are discussed below.
Error summary
Table 9
summarizes Model 2700 calibration errors.