![Keithley 7706 Скачать руководство пользователя страница 41](http://html2.mh-extra.com/html/keithley/7706/7706_user-manual_3982827041.webp)
37
Calibration commands
Table 8
Model 7706 calibration commands
Command Description
:CALibration
Calibration root command.
:PROTected
All commands in this subsystem are protected by the calibration lock (except
queries and :CODE).
:CODE '<up to 8 char. string>'
Send calibration code. (Default KI002700)
:CARD1
Path to card in slot 1.
:INITiate
Initiate card calibration.
:COUNt?
Request number of times card had been calibrated.
:RCOunt
Reset card calibration count to 0.
:DATE?
Request card calibration date.
:SAVE
Save cal constants to card EEPROM.
:LOCK
Lock out calibration.
:LOCK?
Request cal lock state. (0 = locked, 1 = unlocked)
:STEP0 <NRf>
Temperature sensor cold calibration (<NRf> = temperature, °C).
:DAC
Path to analog output calibration.
:STEP0 <NRf>
Analog output DAC calibration.
:CARD2
Path to card in slot 2.
:COUNt?
Request number of times card has been calibrated.
:DATE?
Request card calibration date.
:DIAGnostic Diagnostic
root.
:KEIThley
Keithley path.
:CALibration
Calibration path.
:UNLOCK
Unlock command.