![Keithley 7706 Скачать руководство пользователя страница 31](http://html2.mh-extra.com/html/keithley/7706/7706_user-manual_3982827031.webp)
27
Bus commands
Bus commands
When issuing commands over the bus, keep the following channel usage in mind:
Sending channel numbers
Express channel numbers as a three-digit number where the first digit is the slot number. For example, Channel 1 on slot 1 would
be Channel 101, and Channel 5 on slot 2 would be 205. The hundreds digit is the slot number, and the remaining digits denote
the channel number on the card (see “Channel list parameter” located at the bottom of
Table 4
).
Table 3
Channel description
Channels
Description
1-20
Voltage, resistance, temperature, frequency, period, etc., measurements
21-22
Eight-bit digital outputs
23-24
Sixteen-bit analog outputs
25
Totalizer
26
2/4 pole selection relay. Closing Channel 26 makes a 4-pole measurement
27
Sense terminals to backplane isolation
28
Input terminals to backplane isolation