Model 7706 module
Remote Model 7706 analog output (DAC) calibration
1. Connect the Model 7706 CH23 and 24 to CH1 and 2 (use
the same connection setup as in
Figure 24
)
:
• Connect CH23 H and L terminals to CH1 H and L terminals (H to H; L to L).
• Connect CH24 H and L terminals to CH2 H and L terminals (H to H; L to L).
2. With the power off, install the Model 7706 in Slot 1, and select the rear inputs with the INPUTS switch.
3. Turn on the Model 2700 power.
4. Allow unit to warm up for two hours before proceeding.
5. Unlock calibration by sending:
:DIAG:KEIT:CAL:UNLOCK
6. Enable calibration by sending the :CODE command. For example, the default command is:
:CAL:PROT:CODE 'KI002700'
7. Initiate calibration by sending the following command:
:CAL:PROT:CARD1:INIT
8. Calibrate the Model 7706 analog output with the following command:
:CAL:PROT:CARD1:DAC:STEP0
9. Send the following commands to save calibration, and lock out calibration:
:CAL:PROT:CARD1:SAVE
:CAL:PROT:CARD1:LOCK
Calibration commands
Table 8
summarizes calibration commands for the Model 7706 plug-in module. Note that CARD1 commands calibrate the card
in Slot 1, while CARD2 commands request calibration count and date information from a card in Slot 2.
NOTE
The 7706 must be installed in Slot 1 through a Model 7798 extender card to perform temperature calibration.