Service Information
5-5
Figure 5-2
Differential offset current test connections
Bank Outputs
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
!
11
5
V
!
90-110V
105-125V
180-220V
210-250V
M
odel 7035
Model 6517A
(Measure Current)
HI (Red)
Note:
Setup shown is configured
to test Bank A pathways for
offset current.
Model 7025
Unterminated
Triax Cable
LO (Black)
A) Differential
Bank Inputs
H
L
H
L
H
L
H
L
A
1
3
4
2
B
C
D
E
F
G
H
I
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
Bank Outputs
B) Common-Mode
Bank Inputs
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
!
1
15V
!
90-110V
105-125V
180-220V
210-250V
Model 6517A
(Measure Current)
HI (Red)
Note:
Setup shown is configured
to test Bank A pathways for
offset current.
Model 7025
Unterminated
Triax Cable
Short
Model 7035
LO
(Black)
H
L
H
L
H
L
H
L
A
1
3
4
2
B
C
D
E
F
G
H
I
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
Содержание 7035