7-14
Wave Functions (6221 Only)
Model 6220/6221 User’s Manual
Return to
Programming arbitrary waveforms
1.
Restore defaults with this command
:
*RST
2.
Configure the waveform
.
Examples
– The following commands configure a 100kHz arbitrary wave
with an amplitude of 25mA, 0mA offset, phase marker off, and 20 second
duration:
3.
Arm and trigger the waveform, turn on output:
4.
To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:ARB:DATA -0.5, -0.25,
0, 0.3, 0.4
‘
‘
Define arbitrary data points in
location 0.
SOUR:WAVE:ARB:COPY 1
‘ Copy points to location 1.
SOUR:WAVE:FUNC ARB1
‘ Select arb wave, location 1.
SOUR:WAVE:FREQ 1e5
‘ Set frequency to 100kHz.
SOUR:WAVE:AMPL 25e-3
‘ Set amplitude to 25mA.
SOUR:WAVE:OFFS 0
‘ Set offset to 0.
SOUR:WAVE:PMAR:STAT OFF
‘ Turn off phase marker.
SOUR:WAVE:DUR:TIME 20
20 second duration.
SOUR:WAVE:RANG BEST
‘ Select best fixed source range.
SOUR:WAVE:ARM
‘ Arm waveform.
SOUR:WAVE:INIT
‘ Turn on output, trigger waveform.
SOUR:WAVE:ABOR
‘ Stop generating waveform.
Содержание 6220 DC
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 14: ...7 Wave Functions 6221 Only Figure 7 1 Offset example 7 4 Figure 7 2 Duty cycle 7 5 Figure 7 3 Phase marker 7 6...
Страница 16: ......
Страница 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Страница 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Страница 148: ...A Specifications...
Страница 167: ......
Страница 169: ......
Страница 170: ......