Model 6220/6221 User’s Manual
Delta, Pulse Delta, and Differential Conductance
5-31
Return to
The following query command is used to return the response to a query
command sent over the serial port:
SYSTem:COMMunicate:SERial:ENTer?
When communicating over the serial port, there are no errors reported if a
Model 2182A is not connected to the serial port.
Examples
– The following commands demonstrate proper syntax for
sending commands and returning responses to queries over the serial port:
4.
Set measurement units
– Volts are the default units for the Model 6221
but can instead be expressed (and displayed) as an Ohms, Watts, or Sie-
mens reading (see “
for details). The
commands to set measurement units are listed in
. The following
example shows the command sequence to select average power measure-
ment units:
5.
Set up, arm, and run Pulse Delta
– Details on the commands to set up
and arm Pulse Delta are provided in
. The following example dem-
onstrates the proper sequence to set up, arm, and run a Fixed output:
A.
Trace points specifies the size of the buffer. Buffer size should be the
same value as Pulse Delta count. See
for details on all buffer
commands.
SYST:COMM:SER:SEND “VOLT:RANG 1”
‘ Select 1V range for 2182A.
SYST:COMM:SER:SEND “VOLT:RANG?”
‘ Send range query.
SYST:COMM:SER:ENT?
‘ Return response to query.
UNIT W
‘ Select power (Watts) measurement units.
UNIT:POWer AVERage
‘ Select Average power type.
*RST
‘ Restores 6221 defaults.
SOUR:PDEL:HIGH 10e-3
‘ Sets pulse high value to 10mA.
SOUR:PDEL:LOW 0
‘ Sets pulse low value to 0mA.
SOUR:PDEL:WIDT 500e-6
‘ Sets pulse width to 500µs.
SOUR:PDEL:SDEL 100e-6
‘ Sets source delay to 100µs.
SOUR:PDEL:COUN 200
‘ Sets pulse count to 200.
SOUR:PDEL:RANG BEST
‘ Selects the best source range.
SOUR:PDEL:INT 10
‘ Sets pulse interval to 10 PLC.
SOUR:PDEL:SWE OFF
‘ Disables sweep function.
SOUR:PDEL:LME 2
‘ Set for two low pulse measurements.
TRAC:POIN 200
‘ Sets buffer to 200 points.
A
SOUR:PDEL:ARM
‘ Arms Pulse Delta.
INIT:IMM
‘ Starts Pulse Delta measurements.
B
Содержание 6220 DC
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 14: ...7 Wave Functions 6221 Only Figure 7 1 Offset example 7 4 Figure 7 2 Duty cycle 7 5 Figure 7 3 Phase marker 7 6...
Страница 16: ......
Страница 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Страница 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Страница 148: ...A Specifications...
Страница 167: ......
Страница 169: ......
Страница 170: ......