manualshive.com logo in svg
background image

Model 4225-RPM and Model 4225-RPM-LR Remote Pulse Module Instructions 

 

PA-1086 Rev. A October 2020 

 

Differences between the 4225-RPM and 4225-RPM-LR 

The main difference between the 4225-RPM and the 4225-RPM-LR is output resistance. The 4225-RPM has 
between 20 Ω and 50 Ω of output resistance. The 4225-RPM-LR has 2 Ω to 3 Ω of output resistance. This low 
output resistance is important for customers who need more accurate low voltage output at currents up to 10 

mA without using the Load Line Effect Compensation (LLEC) feature to improve this low voltage accuracy at 

higher currents. 

Another difference between the models is that the 4225-RPM-LR can only drive capacitive loads up to 75 pF. 

This amount includes both the interconnect and the DUT capacitances. 

In addition, the 4225-RPM-LR does not include a hardware current limit. The RPM hardware current limit is a 

feature available only in UTMs. Some users, especially those testing nonvolatile memory, use a hardware 

current limit to limit the maximum current that the RPM outputs. If you need this feature, you should use the 

4225-RPM. Attempts to use this feature on the 4225 RPM-LR generate an error. 

For additional information, including information for users trying to drive high capacitive loads (>75 pF), see 

Load line effect

 (on page 8). 

 

Load-line effect 

The load-line effect causes voltage across the DUT to be lower than the programmed source voltage. The 

output resistance of a pulsing instrument causes this effect. This resistance causes a reduction in output 

voltage at the test device when current flows. By design, pulse instruments such as the 4225-PMU typically 

have 50 

Ω of resistance when placed in series with the output. The resistance causes a voltage drop 

proportional to the current flow and is called the load-line effect. 

The following figure shows the result of the load-line effect when measuring standard transistor V

DS

-I

DS

 curves 

with the 4225-PMU on its 200 mA range. Even though the user specified a source voltage of 6 V with a current 

of about 25 mA (the blue curve at around 25E-3, I

DS

), about 4.8 V is applied to the DUT, while about 1.2 V 

(marked A) is dropped inside the 4225-PMU and is not applied to the DUT. The difference between the actual 

voltage sourced at the load line and the sourced voltage is caused by the load-line effect. The amount shown at 

A is more than 1 V of loss. Increasing the amperage of the curve increases this effect. The amount shown at B 

is more than 2 V of loss. 

The load line is defined by V

SRC

 and the resistance of the pulse instrument. The output resistance of the 

4225-PMU entirely determines its slope. The 4225-PMU programmed source voltage determines its horizontal 

position. Larger voltages shift the line to the right. Smaller voltages shift it to the left. For a given programmed 

source voltage, the actual DUT voltage cannot go beyond this line. 

The slope is defined by the equation: 

slope

 

= −1/R

SRC

 

Содержание 4225-RPM

Страница 1: ...hat automatically switches between precision dc source measure units SMUs C V and the ultra fast pulsed I V instruments In addition the Remot Preamplifier Module RPM extends the low current measuremen...

Страница 2: ...ve cap CAP 47 2 6 Triaxial caps for shielding not shorting CA 31 4 Not shown Support bracket 4200 322 1 Not shown Captive panel screw for support bracket FA 243 2 1 Not shown Allen socket head capscre...

Страница 3: ...PMU and used by the PMU as well as the RPM modules Multi measurement Prober Cable Kit documents These are guides for the optional 4210 MMPC kits that are customized for various probers Switching and c...

Страница 4: ...problems that may affect proper ranging and LLEC operation do not mix 4225 RPMs and 4225 RPM LRs in one system Use either all 4225 RPMS or all 4425 RPM LRs Connections from the 4225 PMU to an RPM The...

Страница 5: ...ions PA 1086 Rev A October 2020 5 Connections to a 4 terminal device The following illustrates the connection to a 4 terminal device using the cabling and adapters that are included with either RPM or...

Страница 6: ...PA 1086 Rev A October 2020 Typical 2 terminal connections The following figure illustrates the connections to 2 terminal devices using the optional MMPC cabling This figure shows two 4225 RPMs or 4225...

Страница 7: ...225 RPM LR extends the lower current ranges of the 225 PMU This is especially important for devices like diodes or transistors that have I V curves that extend over several decades of current The auto...

Страница 8: ...to be lower than the programmed source voltage The output resistance of a pulsing instrument causes this effect This resistance causes a reduction in output voltage at the test device when current fl...

Страница 9: ...put but this is typically between 2 and 3 not including cable resistance It also exhibits the load line effect but to a much lesser degree than the 4225 RPM The following figure compares the load line...

Страница 10: ...n these current ranges the RPM acts as a preamplifier for pulse current measurements Above the 10 mA range the RPM is bypassed and the current measurement occurs in the 4225 PMU which has 50 of output...

Страница 11: ...may be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument...

Страница 12: ...cumentation explains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result i...

Отзывы: