Model 4225-RPM and Model 4225-RPM-LR Remote Pulse Module Instructions 

 

 

PA-1086 Rev. A October 2020 

 

Indicators 

When the RPM is operational in a 4200A-SCS system, an LED on top of the RPM illuminates. The color of the 

LED indicates which instrument is connected to the DUT during a test. Use the color codes in the following 

figure to determine which instrument is connected. Note that the instrument indicator does not show that a test 

is running or that the output of an instrument is active. To determine if a test is running, check the status in 

Clarius. 

Figure 3: Instrument indicator 

 

 

Related documents 

The following documents are available in the 4200A-SCS Learning Center: 

 

Model 4200-PRB-C, 4200 Pulse SMA to SSMC Adapter (PA-928): This document provides information on 

the 4200-PRB-C cable that is provided with the 4225-PMU and used by the PMU as well as the RPM 

modules. 

 

Multi-measurement Prober Cable Kit documents: These are guides for the optional 4210-MMPC kits that 

are customized for various probers. 

 

Switching and connections 

Both modules allow for automatic switching between dc I-V, C-V, and pulsed I-V measurements when executed 

in ITM tests. The following topics describe typical connections for these measurements. 

When properly installed, you can perform different electrical measurements on the device without having to 

disconnect and reconnect cabling for each test. Optional multi-measurement performance cable kits 

(Model 4210-MMPC) are available to connect the 4200A-SCS to a prober manipulator. In addition to eliminating 

the need for re-cabling, these kits help maximize signal fidelity by eliminating the measurement errors that often 

result from cabling errors. 

Turn off system power before connecting or disconnecting the RPM to or from the PMU. 

Failure to do so may result in PMU or RPM damage, possibly voiding the warranty. 

Содержание 4225-RPM

Страница 1: ...hat automatically switches between precision dc source measure units SMUs C V and the ultra fast pulsed I V instruments In addition the Remot Preamplifier Module RPM extends the low current measuremen...

Страница 2: ...ve cap CAP 47 2 6 Triaxial caps for shielding not shorting CA 31 4 Not shown Support bracket 4200 322 1 Not shown Captive panel screw for support bracket FA 243 2 1 Not shown Allen socket head capscre...

Страница 3: ...PMU and used by the PMU as well as the RPM modules Multi measurement Prober Cable Kit documents These are guides for the optional 4210 MMPC kits that are customized for various probers Switching and c...

Страница 4: ...problems that may affect proper ranging and LLEC operation do not mix 4225 RPMs and 4225 RPM LRs in one system Use either all 4225 RPMS or all 4425 RPM LRs Connections from the 4225 PMU to an RPM The...

Страница 5: ...ions PA 1086 Rev A October 2020 5 Connections to a 4 terminal device The following illustrates the connection to a 4 terminal device using the cabling and adapters that are included with either RPM or...

Страница 6: ...PA 1086 Rev A October 2020 Typical 2 terminal connections The following figure illustrates the connections to 2 terminal devices using the optional MMPC cabling This figure shows two 4225 RPMs or 4225...

Страница 7: ...225 RPM LR extends the lower current ranges of the 225 PMU This is especially important for devices like diodes or transistors that have I V curves that extend over several decades of current The auto...

Страница 8: ...to be lower than the programmed source voltage The output resistance of a pulsing instrument causes this effect This resistance causes a reduction in output voltage at the test device when current fl...

Страница 9: ...put but this is typically between 2 and 3 not including cable resistance It also exhibits the load line effect but to a much lesser degree than the 4225 RPM The following figure compares the load line...

Страница 10: ...n these current ranges the RPM acts as a preamplifier for pulse current measurements Above the 10 mA range the RPM is bypassed and the current measurement occurs in the 4225 PMU which has 50 of output...

Страница 11: ...may be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument...

Страница 12: ...cumentation explains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result i...

Отзывы: