4200-900-01 Rev. K / February 2017
Return to
1-17
Model 4200-SCS User’s Manual
Section 1: Getting Started
Most common causes of ABB not locked are:
• Mismatched physical cable lengths
• Mismatched physical cable lengths versus the programmed cable length in Keithley
Interactive Test Environment (KITE)
• Improperly torqued SMA cables
• Sub-optimal I-range setting
• Too much parasitic load on the low side of DUT
Figure 1-12
Measurement circuit (simplified)
The capacitive impedance (and conductance) are calculated based on the measured AC
impedance and phase.
The capacitance is calculated from the capacitive impedance and the test frequency using the
following formula:
Measurement functions
The 4210-CVU can measure the following parameters:
• Z, Theta
Impedance and Phase Angle
• R + jX
Resistance and Reactance
• Cp-Gp
Parallel Capacitance and Conductance
• Cs-Rs
Series Capacitance and Conductance
• Cp-D
Parallel Capacitance and Dissipation Factor
• Cs-D
Series Capacitance and Dissipation Factor
shows the vector diagram and fundamental equations for impedance.
AC
Test
Signal
DC
Bias
4200-CVU
HCUR
Measure
AC Current
HPOT
Measure
AC Voltage
V
LPOT
A
LCUR
DUT
ABB Loop
I
DUT
Red SMA
Cable (1 of 4)*
Shields of the cables must
be connected together near
the DUT.
*
C
DUT
I
DUT
2
π
fV
AC
--------------------
=
C
DUT
= Capacitance of the DUT (f)
f
= Test frequency (Hz)
V
AC
= Measured AC voltage (V)