4200-900-01 Rev. K / February 2017
Return to
3-75
Model 4200-SCS User’s Manual
Section 3: Common Device Characterization Tests
NOTE
To copy between sites, refer to
Clear, copy, paste, and paste to all sites on
NOTE
After setting the device stress properties for all devices and sites, you must do steps
13 and 14 of the above procedure to apply and save the new settings. Failure to do
so will cause the new settings to be lost.
NOTE
In the Device Stress Properties windows, the names for device terminals (for
example, Drain, Gate, Source and Bulk) and the enabled fields for those terminals
are set automatically by KITE. The terminal names correspond to the terminal names
used by the ITMs for the device. When you double-click an ITM in the project
navigator for the device, it will show the schematic of the device and the names of the
terminals.
Setting AC stress properties
With the
AC voltage stress
type selected, DC and AC stress properties can be
set.
shows an example for a device that is connected directly (no
matrix) to channel 2 of the VPU (AC stress). The device is also connected to
SMU1 for DC stress.
•
Device Pin / SMU connections
:
In this area of the window, the 2 pin assignment and the checked box for
VPU
indicates that the gate is connected to channel 2 of the VPU.
7
Use the lists to control Stress Measurements. Options include:
• Do Not Measure
• First Stress Only
• Every Stress Cycle
See
8
For a multi-device Subsite Plan, click
Next Device
to display the stress properties window for the next device. Click
Prev Device
to select the previous device. See
Clear, copy, paste, and paste to all sites
.
9
Repeat steps 2 through 7 for all devices in the Subsite Plan.
10
Repeat steps 1 through 9 to configure the Device Stress Properties for another site.
11
When active, enable to leave the outputs on after the end of a stress cycle. See
12
Click to check SMU or VPU resources and connections for the active (displayed) site. Resource Status will indicate if
enough SMUs and VPUs are present. Connections status will report any problems with matrix connections. See
13
When finished setting the stress properties for all devices and all sites, click
OK
. To cancel all changes and close the
window, click
Cancel
.
14
), click
Apply
to apply and save the settings made in the Device Stress
Properties window.
Table 3-25
Setup steps for the first device in the Subsite Plan (continued)
Step