3-84
Return to
4200-900-01 Rev. K / February 2017
Section 3: Common Device Characterization Tests
Model 4200-SCS User’s Manual
Figure 3-66
Example of “First Stress Only” measurement
Device selection
The Device Stress Properties window corresponds to the selected device in the
Subsite Plan. The individual properties window for the each device is selected
using the
Next Device
or
Prev
Device
buttons. If there is only one device in the
Subsite Plan, these buttons will be disabled.
Leave stress conditions on
DC stressing only
: Enable the
Leave Stress Conditions On
button to leave the
outputs of the SMUs on after the end of a stress cycle.This allows the stress to
continue until the next test is performed in the project tree. You may want to keep
stress on as long as possible so the DUT doesn’t have time to relax before the
tests are performed.
Clear, copy, paste, and paste to all sites
Clear
: Clicking the
Clear
button clears all stress properties data for the displayed
device. It sets all voltage and current values to zero, sets device pin number
assignments to zero, sets Stress Measurements to Do Not Measure, and disables
all Targets (clears Target Values).
Copy
and
Paste
: Copy and Paste allow properties settings for one device to be
copied and pasted into the properties window for a different device. It can also be
used to copy and paste settings into a different site.
41.8017E
-
10
10.00
1
Drop-down menus to control Stress Measurements.
Do Not Measure: Do not Make the specified
measurement.
First Stress Only: Take the specified measurement
on the first stress cycle only.
Every Stress Cycle: Take the specified measurement
on every stress cycle.
Example of First Stress Only measurement (I Drain)
Содержание 4200-SCS
Страница 2: ......
Страница 4: ......
Страница 334: ...Index 6 4200 900 01 Rev K February 2017 Model 4200 SCS User s Manual Index...