Model 3762 10-Channel High Voltage Multiplexer Card User's Manual 

Section 3: General operation 

 

3762-900-01 Rev. A / July 2017 

3-3 

 

Wiring 

Because of the high impedance of the board, take special care when handling and using to prevent 
degradation of performance. Handle the board by the edges to avoid contaminating it with dirt or 
body oil. 

 

 

Each channel on the 3762 consists of a double-pole, normally open (2 Form A) pair of single-pole 
switching relays. The 3762 will switch any one of the 10 signals (inputs) to one output, or switch one 
signal to any of one of 10 outputs. 

 

 

Hazardous voltages may be present on all input, output, and guard terminals. To prevent 

electric shock that could cause injury or death, never make or break connections to the 
3762 while electrical equipment is powered on. Turn off the equipment from the front panel 

or disconnect the main power cord from the rear of the 3762 before handling cables. Turn 
off or disconnect any external equipment from the card before removing the card from the 

mainframe or changing any connections. Precautions must be taken to prevent a shock 
hazard by surrounding the test device and any unprotected leads (wiring) with double 

insulation for 1000 volts, Category I. 

 

Wiring is accomplished by means of terminal strips as shown on the component layout (see 

Wire 

selection and preparation

 (on page 3-2)). Each channel has a HI connection, LO connection, and 

GUARD connection. Guard is common to all channels with the jumpers (R201-R210) installed. The 
3762 is shipped with all jumpers installed.  

1.  Wiring is accomplished by means of terminal strips on screw terminals. 

2. 

Resistance of the relay contacts (terminals) path is less than 0.2 Ω. 

3.  A common guard surrounds all analog signal paths. 

4.  Use wires or cables that are rated for maximum signal levels. The maximum allowable wire size 

is No. 16 AWG and the minimum size is No. 26 AWG. 

When operating at high current levels (amperes) consider the voltage burden (I x R drop) of the wire 
or cable that is used. For example, the approximate resistance of 24 gauge copper wire is 26 

mΩ/ft. 

The voltage burden is approximately 26 mV/ft/amp. At the maximum signal current of 1 ampere the 
voltage burden would be 26 mV/ft. If the wire being used is 4 feet long, the voltage burden of the wire 
is approximately 0.1 V. 

5.  Route the wires through the rubber clamp pads located at the rear of the card. 

 

Installation and removal 

Once the card is wired, insert it card-edge first into the system switch mainframe by aligning it with the 
grooves in the appropriate slot. Make sure it is properly seated into the mainframe connector. 

To remove a card, first turn off the mainframe and all other equipment connected to the card. Grasp 
the end of the card and carefully pull it out of the mainframe. 

 

 

Содержание 3706A

Страница 1: ...User s Manual 3762 900 01 Rev A July 2017 P3762 900 01A 3762 900 01A www tek com keithley Model 3762 10 Channel High Voltage Multiplexer Card...

Страница 2: ...nts is strictly prohibited These are the original instructions in English All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments Other brand names are t...

Страница 3: ...be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to...

Страница 4: ...damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or minor injury or damage th...

Страница 5: ...lation 2 2 General operation 3 1 General operation 3 1 Safety hazards 3 1 Wire selection and preparation 3 2 Wiring 3 3 Installation and removal 3 3 Operation considerations 4 1 Cables 4 1 External fa...

Страница 6: ...Table of contents Model 3762 10 Channel High Voltage Multiplexer Card User s Manual Index I 1...

Страница 7: ...for shielding or as a guard driven from a single instrument Guards may be isolated by removing jumpers installed at each input Multiple switched guard circuits can be achieved by removing the jumper a...

Страница 8: ...are requirements switching card installation instructions and a brief description of front panel and remote interface operation 3700AS 900 01 Series 3700A System Switch Multimeter User s Manual Inform...

Страница 9: ...l with a user selectable shield or driven guard connection The output connection and the input connections have screw terminals with a maximum allowable wire size No 16 The schematic representation is...

Страница 10: ...u are qualified Failure to recognize and observe standard safety precautions could result in personal injury or death due to electric shock Unpack and inspect The 3762 was inspected both electrically...

Страница 11: ...are provided internal to the card Ensure that any signals connected to the multimeter or mainframe are limited to the allowable ratings of the device For more information about wire connections see Wi...

Страница 12: ...gn the card s edge into the slot s card guide and slide in the card For the last inch or so press in firmly to mate the card connector to the mainframe connector 5 On each side of the card there is a...

Страница 13: ...Section 2 Installation and connections Model 3762 10 Channel High Voltage Multiplexer Card User s Manual 2 4 3762 900 01 Rev A July 2017...

Страница 14: ...k connections while the equipment is powered on User supplied lethal voltages may be present on the PC board or the connections Turn off the equipment from the front panel disconnect the main power co...

Страница 15: ...ounting screws for the card are securely connected to the system switch mainframe chassis in order to provide the proper chassis protective grounding 9 Read this manual carefully and adhere to all of...

Страница 16: ...on the component layout see Wire selection and preparation on page 3 2 Each channel has a HI connection LO connection and GUARD connection Guard is common to all channels with the jumpers R201 R210 in...

Страница 17: ...interfering with any equipment The cable rating must not exceed the maximum allowable signal levels as defined in the specifications External factors The 3762 is specified for a resistive load however...

Страница 18: ...ve load must be less than 0 5 A to protect the relays The next figure shows typical circuits to limit current surges Also consider the maximum load of 10 VA when determining the current limit For exam...

Страница 19: ...in the next figure would be R V I 250 V 40 mA 6 25 k Figure 7 Model 3762 Limiting inductive reaction voltage Make a modification The guard for channels on the 3762 high voltage multiplexer card are al...

Страница 20: ...the testing of multiple devices for more information about switching go to the Keithley Instruments website http www tek com keithley and search for Switching Handbook Shielded application This appli...

Страница 21: ...switched to a DUT using a driven guard as shown in the next figure the guard must be switched The jumper for each channel in use is removed so that the various sources will not try to drive the guards...

Страница 22: ...acy requirements Figure 11 3762 Source with external sense or Kelvin Measurement HI and Sense HI S are switched through one card The jumper is installed to provide a guard all the way to the DUT Since...

Страница 23: ...re A double pole switch normally open is classified as a 2 Form A Form B is similar to Form A except that its contacts are normally closed see next figure A double pole switch normally closed is class...

Страница 24: ...nstallation and connections 2 1 Installation and removal 3 3 Introduction 1 1 M Make a modification 4 3 Model 3762 typical applications 5 1 modules installation 2 2 O Operation considerations 4 1 R Re...

Страница 25: ...names are the property of Keithley Instruments All other trademarks and trade names are the property of their respective companies Keithley Instruments Corporate Headquarters 28775 Aurora Road Clevel...

Отзывы: