Model 3762 10-Channel High Voltage Multiplexer Card User's Manual 

Section 2: Installation and connections 

 

3762-900-01 Rev. A / July 2017 

2-3 

 

Perform the following steps to install a switching card into the instrument mainframe: 

1.  Turn the instrument off and disconnect the power line cord and any other cables connected to the 

rear panel. 

2.  Position the instrument so that you are facing the rear panel. 

3.  Remove the slot cover plate from the desired mainframe slot. Retain the plate and screws for 

future use. 

4.  With the top cover of the switching card facing up, align the card's edge into the slot's card guide 

and slide in the card. For the last ¼ inch or so, press in firmly to mate the card connector to the 
mainframe connector. 

5.  On each side of the card, there is a mounting screw. Tighten these two screws to secure the card 

to the mainframe. Do not overtighten. 

6.  Reconnect the power line cable and any other cables to the rear panel, then turn the instrument 

on. 

7.  Press the

 SLOT

 key to see the model numbers, description, and the firmware revision of the 

installed switching cards, along with the mainframe firmware and DMM (if present). 

Item shipped may vary from model pictured here. 

 

Figure 3: Typical module installation 

 

 
 

 

Item

 

Description

 

Card guide (part of mainframe) 

Card 

Card edge (part of card) 

Mounting screw (part of card) 

 
 
 
 

 

Содержание 3706A

Страница 1: ...User s Manual 3762 900 01 Rev A July 2017 P3762 900 01A 3762 900 01A www tek com keithley Model 3762 10 Channel High Voltage Multiplexer Card...

Страница 2: ...nts is strictly prohibited These are the original instructions in English All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments Other brand names are t...

Страница 3: ...be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to...

Страница 4: ...damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or minor injury or damage th...

Страница 5: ...lation 2 2 General operation 3 1 General operation 3 1 Safety hazards 3 1 Wire selection and preparation 3 2 Wiring 3 3 Installation and removal 3 3 Operation considerations 4 1 Cables 4 1 External fa...

Страница 6: ...Table of contents Model 3762 10 Channel High Voltage Multiplexer Card User s Manual Index I 1...

Страница 7: ...for shielding or as a guard driven from a single instrument Guards may be isolated by removing jumpers installed at each input Multiple switched guard circuits can be achieved by removing the jumper a...

Страница 8: ...are requirements switching card installation instructions and a brief description of front panel and remote interface operation 3700AS 900 01 Series 3700A System Switch Multimeter User s Manual Inform...

Страница 9: ...l with a user selectable shield or driven guard connection The output connection and the input connections have screw terminals with a maximum allowable wire size No 16 The schematic representation is...

Страница 10: ...u are qualified Failure to recognize and observe standard safety precautions could result in personal injury or death due to electric shock Unpack and inspect The 3762 was inspected both electrically...

Страница 11: ...are provided internal to the card Ensure that any signals connected to the multimeter or mainframe are limited to the allowable ratings of the device For more information about wire connections see Wi...

Страница 12: ...gn the card s edge into the slot s card guide and slide in the card For the last inch or so press in firmly to mate the card connector to the mainframe connector 5 On each side of the card there is a...

Страница 13: ...Section 2 Installation and connections Model 3762 10 Channel High Voltage Multiplexer Card User s Manual 2 4 3762 900 01 Rev A July 2017...

Страница 14: ...k connections while the equipment is powered on User supplied lethal voltages may be present on the PC board or the connections Turn off the equipment from the front panel disconnect the main power co...

Страница 15: ...ounting screws for the card are securely connected to the system switch mainframe chassis in order to provide the proper chassis protective grounding 9 Read this manual carefully and adhere to all of...

Страница 16: ...on the component layout see Wire selection and preparation on page 3 2 Each channel has a HI connection LO connection and GUARD connection Guard is common to all channels with the jumpers R201 R210 in...

Страница 17: ...interfering with any equipment The cable rating must not exceed the maximum allowable signal levels as defined in the specifications External factors The 3762 is specified for a resistive load however...

Страница 18: ...ve load must be less than 0 5 A to protect the relays The next figure shows typical circuits to limit current surges Also consider the maximum load of 10 VA when determining the current limit For exam...

Страница 19: ...in the next figure would be R V I 250 V 40 mA 6 25 k Figure 7 Model 3762 Limiting inductive reaction voltage Make a modification The guard for channels on the 3762 high voltage multiplexer card are al...

Страница 20: ...the testing of multiple devices for more information about switching go to the Keithley Instruments website http www tek com keithley and search for Switching Handbook Shielded application This appli...

Страница 21: ...switched to a DUT using a driven guard as shown in the next figure the guard must be switched The jumper for each channel in use is removed so that the various sources will not try to drive the guards...

Страница 22: ...acy requirements Figure 11 3762 Source with external sense or Kelvin Measurement HI and Sense HI S are switched through one card The jumper is installed to provide a guard all the way to the DUT Since...

Страница 23: ...re A double pole switch normally open is classified as a 2 Form A Form B is similar to Form A except that its contacts are normally closed see next figure A double pole switch normally closed is class...

Страница 24: ...nstallation and connections 2 1 Installation and removal 3 3 Introduction 1 1 M Make a modification 4 3 Model 3762 typical applications 5 1 modules installation 2 2 O Operation considerations 4 1 R Re...

Страница 25: ...names are the property of Keithley Instruments All other trademarks and trade names are the property of their respective companies Keithley Instruments Corporate Headquarters 28775 Aurora Road Clevel...

Отзывы: