Section 3: Function and features
Series 2230G Triple-Channel Power Supplies User's Manual
3-6
2230G-900-01 Rev. A / June 2018
Set the current output or current limit for a specific channel
You can set the current limit from 0 A to the maximum current specified for your Series 2230G.
To set the current limit:
1. Select the channel by pressing the
CH1
,
CH2
, or
CH3
key.
2. Press the
I-Set
key.
3. Use the numeric keys, the navigation wheel, or the up, down, right, and left arrow keys to set the
current limit.
4. Press
Enter
to confirm the setting.
Save and recall setups
You can store up to 36 different setups in Series 2230G memory. Each setup includes a voltage limit,
current limit, and maximum output voltage for each channel.
To save the setups:
1. After you set up the Series 2230G, press the
Save
key.
2. Use the number keys, the navigation wheel, or the arrow keys to select the setup memory (1 to
36) that you want to store the values in.
3. Press
Enter
to confirm your setting.
To recall the setups:
1. Press the
Recall
key.
2. Use the number keys, the navigation wheel, or the arrow keys to select the setup that you want to
recall.
3. Press
Enter.
You can also use the following SCPI command to save or recall the setup.
*SAV
*RCL
The Series 2230G does not support the save and recall functions when the channels are in tracking
mode or when they are wired in series or parallel.
Restore default settings
To restore the factory default settings:
1. Remove all of the leads from the output terminals.
2. Press the
Power
key to turn on the power supply.
3. Press
Menu
.
4. Press
Enter
to select
Default Set
.
5. Press the down arrow key to select
Yes
.
6. Press
Enter
. The display shows
Defaults Restored!
. The default settings for the menu
selections are shown in the following table.