Calibration
2-16
26. Press ENTER, and note that the instrument displays the
following during 1M¾ calibration:
Performing 1.0 MOhm Calibration
27. At this point, the instrument will display the following
message advising you to disconnect test leads:
OPEN CIRCUIT INPUTS
28. Disconnect all test leads from the INPUT and SENSE
jacks, then press ENTER. During this calibration phase,
the instrument will display the following:
Performing Open-Ckt Calibration
29. After open circuit calibration, the instrument will dis-
play the following message:
AC CALIBRATION PHASE
30. Make sure all test leads are still disconnected from the
Model 2001 INPUT and SENSE jacks.
31. Press ENTER to perform AC calibration, which will
take a while to complete. During AC calibration, the in-
strument will display the following:
Calibrating AC: Please wait
32. After the AC calibration phase is completed, the instru-
ment will display the following:
AC CAL COMPLETE
33. Press ENTER. The instrument will display the follow-
ing to indicate the start of the low-level calibration
phase:
LOW-LEVEL CAL PHASE
NOTE
Use the exact calibration values shown
when performing the following steps.
34. Connect the calibrator to the INPUT terminals, as
shown in Figure 2-3.
35. Press ENTER. The instrument will display the follow-
ing:
Connect 20V @ 1kHz
36. Set the calibrator to output 20V AC at a frequency of
1kHz, then press ENTER. The instrument will display
the following:
Low-Level Cal - Step 1 of 15
37. Next, the instrument will prompt for a new calibration
signal:
Connect 20V @ 30kHz
38. Program the calibrator for an output voltage of 20V AC
at 30kHz, then press ENTER. The instrument will dis-
play the following while calibrating this step:
Low-Level Cal - Step 2 of 15
39. The Model 2001 will then display:
Connect 200V @ 1kHz
40. Set the calibrator output to 200V AC at a frequency of
1kHz, then press ENTER. The Model 2001 will display
the following message:
Low-Level Cal - Step 3 of 15
41. When finished with this step, the Model 2001 will dis-
play:
Connect 200V @ 30kHz
42. Set the calibrator output to 200V AC at 30kHz, then
press ENTER. The Model 2001 will display the follow-
ing:
Low-Level Cal - Step 4 of 15
Figure 2-3
Calibration voltage connections
NEXT
DISPLAY
PREV
POWER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
STORE
RECALL
INFO
LOCAL
CHAN
SCAN
CONFIG
MENU
EXIT
ENTER
RANGE
AUTO
FILTER
MATH
RANGE
2001 MULTIMETER
SENSE
Ω
4 WIRE
HI
INPUT
LO
INPUTS
CAL
500V
PEAK
F
R
FRONT/REAR
2A 250V
AMPS
350V
PEAK
1100V
PEAK
Input HI
Output HI
Input
LO
Output
LO
Model 2001
5700A Calibrator
Ground link installed.
Note
: Use internal Guard (EX GRD LED is off).