Performance Verification
1-6
NOTE
Do not use REL to null offsets when per-
forming AC volts tests. Also, do not en-
able the filter.
5. Select the low-frequency mode as follows:
A. Press CONFIG ACV, select AC-TYPE, then press
ENTER.
B. Select LOW-FREQ-RMS, then press ENTER.
C. Press EXIT as required to return to normal display.
6. Set the calibrator output to 190.000mVAC at a frequen-
cy of 10Hz, and allow the reading to settle.
7. Verify that the Model 2001 reading is within the limits
summarized in Table 1-4.
8. Repeat steps 6 and 7 for 190mVAC at the remaining fre-
quencies listed in the table.
9. Repeat steps 6 through 8 for the 2V, 20V, 200V, and
750VAC ranges, using the input voltages and limits stat-
ed in Table 1-4.
CAUTION
Do not apply more than 400V at 50kHz,
80V at 250kHz, 40V at 500kHz, or 20V
at 1MHz, or instrument damage may
occur.
Figure 1-2
Connections for AC volts verification (all except 2MHz test)
NEXT
DISPLAY
PREV
POWER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
STORE
RECALL
INFO
LOCAL
CHAN
SCAN
CONFIG
MENU
EXIT
ENTER
RANGE
AUTO
FILTER
MATH
RANGE
2001 MULTIMETER
SENSE
Ω
4 WIRE
HI
INPUT
LO
INPUTS
CAL
500V
PEAK
F
R
FRONT/REAR
2A 250V
AMPS
350V
PEAK
1100V
PEAK
1. 900000 VAC RMS
Input HI
Output HI
Input
LO
Output
LO
Model 2001
5700A Calibrator (Output AC Voltage)
5725 Amplifier (Connect to calibrator)
CA-18-1 Low-
capacitance cable
Ground link installed.
Note
: Use internal Guard (EX GRD LED is off).
10. Set the calibrator output to 190.0000mV at a frequency
of 2MHz.
11. Verify that the reading is within limits stated in Table 1-
3.
12. Repeat steps 10 and 11 for 1.900V input on the 2V
range.
Low-frequency mode
1. Turn on the Model 2001, calibrator, and amplifier, and
allow a one-hour warm-up period before making mea-
surements.
2. Connect the Model 2001 to the calibrator, as shown in
Figure 1-2. Be sure to connect the amplifier HI to Model
2001 INPUT HI, and amplifier LO to Model 2001 IN-
PUT LO as shown. Connect the power amplifier to the
calibrator using the appropriate connector on the rear of
the calibrator.
3. Restore Model 2001 factory default conditions, as ex-
plained in paragraph 1.7.
4. Select the ACV function and the 200mV range on the
Model 2001, and make sure that REL is disabled.