Performance Verification
1-3
Table 1-1 (cont.)
Recommended equipment for performance verification
Mfg.
Model
Description
Specifications*
Fluke
5700A
Calibrator
AC current, 40Hz-10kHz (40Hz-1kHz specifications):
190µA: ±245ppm
1.9mA: ±160ppm
19mA: ±160ppm
190mA: ±170ppm
1.9A: ±670ppm
Resistance:
19¾: ±26ppm
190¾: ±17ppm
1.9k¾: ±12ppm
19k¾: ±11ppm
190k¾: ±13ppm
1.9M¾: ±19ppm
19M¾: ±47ppm
100M¾: ±120ppm
Fluke
5725A
Amplifier
AC voltage, 1kHz-10kHz: 750V: ±85ppm
Fluke
5700A-03
Wideband AC option
190mV, 1.9V @ 2MHz, ±0.1%
Fluke
5440A-7002 Low thermal cable set
Keithley CA-18-1
Low capacitance cable
Low capacitance dual banana to dual banana shielded cable (for
ACV), 1.2m (4 ft.) in length.
Keithley R-289-1G
1G¾ resistor
NOTE: Resistor should be characterized to within ±10,000 ppm and
mounted in shielded test box (see procedure).
Metal component box (for
1G¾ resistor)
Insulated banana plugs (2)
(for test box)
Keithley 3940
Multifunction Synthesizer 1Hz-15MHz, ±5ppm
General
Radio
1433-T
Precision Decade Resis-
tance Box
10-400¾, ±0.02%
⎯
⎯
Megohmmeter
1G¾, ±1%
* 90-day calibrator specifications shown include total uncertainty at specified output. The 1.9V output includes 0.5ppm transfer uncertainty. See Appendix
D for recommendation on alternate calibration sources.