Calibration
2-9
2. Disconnect all test leads from the INPUT and SENSE
jacks, then press ENTER. During this calibration phase,
the instrument will display the following:
Performing Open-Ckt Calibration
Step 8: AC self-calibration
1. After open circuit calibration, the instrument will dis-
play the following message:
AC CALIBRATION PHASE
2. Make sure all test leads are still disconnected from the
Model 2001 INPUT and SENSE jacks.
3. Press ENTER to perform AC calibration, which will
take about six minutes to complete. During AC calibra-
tion, the instrument will display the following:
Calibrating AC: Please wait
4. When AC calibration is finished, the instrument will dis-
play the following:
AC CAL COMPLETE
Step 9: Enter calibration dates
1. Press ENTER, and note that the instrument prompts you
to enter the present calibration date:
CAL DATE: 01/01/92
2. Use the cursor and range keys to enter the current date
as the calibration date, then press ENTER. Press EN-
TER again to confirm the date as being correct.
3. The instrument will then prompt you to enter the due
date for next calibration:
NEXT CAL: 01/01/93
4. Use the cursor and range keys to set the date as desired,
then press ENTER. Press ENTER a second time to con-
firm your selection.
Step 10: Save calibration constants
1. At the end of a successful calibration cycle, the instru-
ment will display the following:
CALIBRATION SUCCESS
2. If you wish to save calibration constants from the proce-
dure just completed, press ENTER.
3. If you do not want to save calibration constants from the
procedure just completed and wish instead to restore
previous constants, press EXIT.
4. Press EXIT to return to normal display after calibration.
NOTE
Comprehensive calibration will be auto-
matically locked out after the calibration
procedure has been completed.
2.8.3 IEEE-488 bus comprehensive calibration
Follow the procedure outlined below to perform comprehen-
sive calibration over the IEEE-488 bus. Use the program list-
ed in paragraph 2.6.3 or other similar program to send
commands to the instrument. Table 2-5 summarizes the cal-
ibration procedure and bus commands.
Procedure
Step 1: Prepare the Model 2001 for calibration
1. Connect the Model 2001 to the IEEE-488 bus of the
computer using a shielded IEEE-488 cable such as the
Keithley Model 7007.
2. Turn on the power, and allow the Model 2001 to warm
up for at least one hour before performing calibration.
3. Unlock calibration by briefly pressing in on the recessed
front panel CAL switch, and verify that the following
message is displayed:
CALIBRATION UNLOCKED
Comprehensive calibration can now be run
NOTE
You can query the instrument for the state
of the comprehensive CAL switch by us-
ing the following query:
:CAL:PROT:SWIT?
A returned value of 1 indicates that cali-
bration is locked, while a returned value of
0 shows that calibration is unlocked.
4. Make sure the primary address of the Model 2001 is the
same as the address specified in the program you will be
using to send commands (see paragraph 2.6.3).