Jazz Ultra Instruments by KEELER

•  The product and the ear specula are supplied non-sterile. Do not use on injured tissue.
•  Use new or sanitised specula to limit the risk of cross-contamination.
•  The disposal of used ear specula must occur in accordance with current medical practices 

or local regulations regarding the disposal of infectious, biological medical waste.

•  Should the unit not be used for an extended period of time or whilst travelling, remove 

batteries from the handles.

•  Insert new batteries when light intensity of the unit is reduced, thus affecting examination.
•  For maximum light yield it is recommended to always insert new high-quality batteries on 

replacement.

•  Never immerse handles in fl uid. Ensure that no fl uid or condensation penetrates into the 

handle.

•  Do not use a battery that is deformed, leaking, corroded or visually damaged. Handle 

a damaged or leaking battery with care. If you come into contact with electrolyte, 

wash exposed area with soap and water. If it contacts the eye, seek medical attention 

immediately.

•  Do not mix battery types.
•  Do not disassemble or modify the battery. There are no serviceable parts inside
•  Dispose of batteries in line with local environmental regulations.
•  Tape over battery contacts to avoid short circuiting during disposal.

•  After removal of the battery do not touch the battery contacts and the patient 

simultaneously.

•  Do not exceed maximum recommended exposure time.

•  After removal of the bulb / LED do not touch the bulb / LED contacts and the 

patient simultaneously.

2.3  CONTRAINDICATION

There is no restriction to patient population this device can be used with other than those 

outlined in the contraindications stated below.
Due to high illumination levels, the Ophthalmoscope can cause some discomfort in 

photophobic patients. 
Mydriatic agents when used in ophthalmoscopy can cause temporary symptoms of blurred 

vision and photophobia. Adverse reactions to mydriatic drops are rare. 
There are very few risks associated with Otoscopy. Some patients may report a slight 

discomfort during the procedure, especially during the insertion of the speculum into a swollen 

and infl amed ear canal. If the plastic tip of the Otoscope is not replaced or cleaned properly, 

infection can spread from one ear to the other.

Содержание Jazz Ultra

Страница 1: ...Jazz Ultra Instruments INSTRUCTIONS FOR USE A world without vision loss...

Страница 2: ......

Страница 3: ...7 5 OTOSCOPE AND ACCESSORIES 7 5 1 INSERTION AND REMOVAL OF EAR SPECULUM 7 5 2 INTRODUCTION OF EXTERNAL INSTRUMENTS INTO THE EAR 7 5 3 REPLACEMENT OF LED OTOSCOPE 8 6 OPHTHALMOSCOPE AND ACCESSORIES 8...

Страница 4: ...Community Humidity limitation Catalogue number Serial number Translation Medical device The Keeler Jazz Ultra Instruments are designed and built in conformity with Directive 93 42 EEC Regulation EU 20...

Страница 5: ...nts When set to a high power and magnification it can also be used to examine the anterior segment of the eye which includes the eyelids cornea sclera conjunctiva iris aqueous crystalline lens and ant...

Страница 6: ...actitioner This device is intended to be used only by suitably trained and authorised healthcare professionals This product should not be immersed in fluid Instrument may become hot if used for extend...

Страница 7: ...diately Do not mix battery types Do not disassemble or modify the battery There are no serviceable parts inside Dispose of batteries in line with local environmental regulations Tape over battery cont...

Страница 8: ...taminated with cerumen ear drainage or blood Dispose of safely 2 Manually clean all surfaces of the units using a suitable brush and de ionized water detergent solution 2 detergent by volume Ensure th...

Страница 9: ...head from the handle in an anti clockwise direction Remove the existing ring and replace with a new ring in the colour of your choice Screw the instrument heads back on in a clockwise direction 5 OTO...

Страница 10: ...Plus values are displayed in black digits and minus values in red digits 6 2 APERTURES AND FILTERS The following apertures and or filters may be selected by the Aperture and Filter Wheel Small Circle...

Страница 11: ...by authorised Keeler representatives Your Keeler product is guaranteed for 3 years and will be replaced or repaired free of charge subject to the following Any fault due to faulty manufacture The ins...

Страница 12: ...e very low and are not likely to cause any interference in nearby electronic equipment Class B Harmonic emissions IEC 61000 3 2 Not Applicable The Jazz Ultra Otoscope Ophthalmoscope is suitable for us...

Страница 13: ...z magnetic field IEC 61000 4 8 30 A m 30 A m Power frequency magnetic fields should be at a level characteristic of a typical location in a typical professional healthcare facility environment Note UT...

Страница 14: ...e location in which the Jazz Ultra Otoscope Ophthalmoscope is used exceeds the applicable RF compliance level above the Jazz Ultra Otoscope Ophthalmoscope should be observed to verify normal operation...

Страница 15: ...HNICAL SPECIFICATIONS Dimensions Otoscope 12cm x 3cm x 7 5cm H x D x W including handle and speculum Ophthalmoscope 13cm x 3cm x 3cm H x D x W including handle Weight Otoscope 68gm including handle wi...

Страница 16: ...P 7108 Jazz 4mm Specula Pack of 100 1514 P 7116 Jazz 5mm Specula Pack of 100 1514 P 7124 12 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the p...

Страница 17: ...15 EN Jazz Ultra Instruments by KEELER...

Страница 18: ......

Страница 19: ......

Страница 20: ...2021 USA Sales Office Keeler USA 3222 Phoenixville Pike Building 50 Malvern PA 19355 USA Toll Free 1 800 523 5620 Tel 1 610 353 4350 Fax 1 610 353 7814 India Office Keeler India Halma India Pvt Ltd P...

Отзывы: