Jazz Ultra Instruments by KEELER

6.3  REPLACEMENT OF LED – OPHTHALMOSCOPE

Remove instrument head from battery handle. The LED 

is located in the bottom section of the instrument head. 

Remove LED from the instrument head by using your 

thumb and forefi nger. Insert a new LED with the pin on the 

LED fi tted in the recess / slot provided on the base of the 

instrument.

Caution: LED may be hot.

7.  MAINTENANCE

Jazz Ultra instruments and their accessories do not require any specifi c maintenance. Should 

an instrument have to be examined for any reason, please return it to your supplier or an 

authorised dealer in your area.

8.  WARRANTY

No user serviceable parts – all preventative maintenance and servicing must only be performed 

by authorised Keeler representatives.
Your Keeler product is guaranteed for 3 years and will be replaced, or repaired free of charge 

subject to the following:

•  Any fault due to faulty manufacture.
•  The instrument and accessories have been used in compliance with these instructions.
•  Proof of purchase accompanies any claim.

Please note:

•  LEDs are guaranteed for 1 year from date of purchase.

The manufacturer declines any and all responsibility and warranty 

coverage should the instrument be tampered with in any manner or 

should routine maintenance be omitted or performed in manners not in 

accordance with these manufacturer’s instructions.
There are no user serviceable parts in this instrument. Any servicing or 

repairs should only be carried out by Keeler Ltd. or by suitably trained and 

authorised distributors. Service manuals will be available to authorised 

Keeler service centres and Keeler trained service personnel.

Содержание Jazz Ultra

Страница 1: ...Jazz Ultra Instruments INSTRUCTIONS FOR USE A world without vision loss...

Страница 2: ......

Страница 3: ...7 5 OTOSCOPE AND ACCESSORIES 7 5 1 INSERTION AND REMOVAL OF EAR SPECULUM 7 5 2 INTRODUCTION OF EXTERNAL INSTRUMENTS INTO THE EAR 7 5 3 REPLACEMENT OF LED OTOSCOPE 8 6 OPHTHALMOSCOPE AND ACCESSORIES 8...

Страница 4: ...Community Humidity limitation Catalogue number Serial number Translation Medical device The Keeler Jazz Ultra Instruments are designed and built in conformity with Directive 93 42 EEC Regulation EU 20...

Страница 5: ...nts When set to a high power and magnification it can also be used to examine the anterior segment of the eye which includes the eyelids cornea sclera conjunctiva iris aqueous crystalline lens and ant...

Страница 6: ...actitioner This device is intended to be used only by suitably trained and authorised healthcare professionals This product should not be immersed in fluid Instrument may become hot if used for extend...

Страница 7: ...diately Do not mix battery types Do not disassemble or modify the battery There are no serviceable parts inside Dispose of batteries in line with local environmental regulations Tape over battery cont...

Страница 8: ...taminated with cerumen ear drainage or blood Dispose of safely 2 Manually clean all surfaces of the units using a suitable brush and de ionized water detergent solution 2 detergent by volume Ensure th...

Страница 9: ...head from the handle in an anti clockwise direction Remove the existing ring and replace with a new ring in the colour of your choice Screw the instrument heads back on in a clockwise direction 5 OTO...

Страница 10: ...Plus values are displayed in black digits and minus values in red digits 6 2 APERTURES AND FILTERS The following apertures and or filters may be selected by the Aperture and Filter Wheel Small Circle...

Страница 11: ...by authorised Keeler representatives Your Keeler product is guaranteed for 3 years and will be replaced or repaired free of charge subject to the following Any fault due to faulty manufacture The ins...

Страница 12: ...e very low and are not likely to cause any interference in nearby electronic equipment Class B Harmonic emissions IEC 61000 3 2 Not Applicable The Jazz Ultra Otoscope Ophthalmoscope is suitable for us...

Страница 13: ...z magnetic field IEC 61000 4 8 30 A m 30 A m Power frequency magnetic fields should be at a level characteristic of a typical location in a typical professional healthcare facility environment Note UT...

Страница 14: ...e location in which the Jazz Ultra Otoscope Ophthalmoscope is used exceeds the applicable RF compliance level above the Jazz Ultra Otoscope Ophthalmoscope should be observed to verify normal operation...

Страница 15: ...HNICAL SPECIFICATIONS Dimensions Otoscope 12cm x 3cm x 7 5cm H x D x W including handle and speculum Ophthalmoscope 13cm x 3cm x 3cm H x D x W including handle Weight Otoscope 68gm including handle wi...

Страница 16: ...P 7108 Jazz 4mm Specula Pack of 100 1514 P 7116 Jazz 5mm Specula Pack of 100 1514 P 7124 12 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipment This symbol on the p...

Страница 17: ...15 EN Jazz Ultra Instruments by KEELER...

Страница 18: ......

Страница 19: ......

Страница 20: ...2021 USA Sales Office Keeler USA 3222 Phoenixville Pike Building 50 Malvern PA 19355 USA Toll Free 1 800 523 5620 Tel 1 610 353 4350 Fax 1 610 353 7814 India Office Keeler India Halma India Pvt Ltd P...

Отзывы: