Recommended separation distances between portable and mobile RF equipment and the Cryomatic
The Cryomatic is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the
Cryomatic can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Cryomatic as recommended below, according to the maximum power output of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
d = 1.2
√
P
80MHz to 800MHz
d = 1.2
√
P
800MHz to 2.5GHz
d = 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by the absorption and reflection from
structures, objects and people.
The Cryomatic should not be used adjacent to or stacked with any other equipment. If this configuration is required then normal operation of the
Cryomatic should be verified in this setting.
Annex I - EMC Statement and Guidelines