Freeze/Defrost Cycles

 

Freezing of the Cryo-Tip probe is controlled manually by the operator 
using the footswitch. 

   

1

  Ensure that the Cryo-Tip probe is correctly    

 

   

  positioned.

 
   

2

  Press the footswitch down. Freezing starts    

 

   

  immediately and the digital timer will increment. 

   

3

  An audible warning sounds every second during the  

   

  freeze cycle and the Freezing Symbol is displayed. 

   

4

  In addition there is a graphical indication of probe   

   

  performance. 

   

5

  De-frosting is achieved by releasing the footswitch.   

   

  The timer will stop counting and the Defrosting  

 

   

  Symbol is displayed.

   

6

  Subsequent freeze cycles can be carried out by simply  

   

  repeating steps 1-4 as soon as the Ready Symbol is    

   

  displayed.

The freezing function is often accompanied by a characteristic ‘pulsing’ 
sound which indicates that the Cryomatic is regulating the gas to 
the optimum pressure for the probe. The ‘pulsing’ may vary or cease 
completely depending on the pressure of gas in the cylinder.

If the probe performance drops below 100% permanently check the 
cylinder gas pressure or suspect a blocked probe.

Cryo-Tip Probe Disconnection 

   

1

  When the procedure is complete the Cryo-Tip probe  

   

  can be disconnected by pressing the probe release    

   

  button on the probe coupling body.

   

2

  When the probe has been disconnected the system   

   

  will automatically check the gas supply in readiness   

   

  for the next use. This will be indicated by an activity  

   

  bar with the Cylinder Symbol.

   

3

  When this short check is complete the system will be  

   

  ready for the connection of another Cryo-Tip probe.

The system shuts off the gas supply immediately the Cryo-Tip probe is 
disconnected providing a safe condition in the event of inadvertent 
operation of the release button.

It is not recommended that the Cryo-Tip probe is disconnected during 
use as this could result in a pressure lock within the probe assembly 
which may make reconnection more difficult.

5. Operation

Содержание Cryomatic

Страница 1: ...Next Home Back 1 Keeler Instructions for use...

Страница 2: ...ons End of Use 6 Cryo Tip Probes 7 Cleaning Sterilisation Cleaning Steam Sterilisation Cleaning the Console 8 Servicing and Preventative Maintenance Scheduled Maintenance User Maintenance 9 Spare Part...

Страница 3: ...unity Relevant accessories are listed in the Spare Parts Accessories section For your own safety and the safety of the equipment always take the following precautions Keep the console away from source...

Страница 4: ...on Dioxide gas to function this is the responsibility of the user Console The Cryomatic console is a self contained system The console provides the connection points for the Cryo Tip probe footswitch...

Страница 5: ...as cylinder status and freeze timer 3 Footswitch Connection Connection point for footswitch plug 4 Cryo Tip Probe Interface Pneumatic and electrical connections for probe 5 Electrical Input IEC connec...

Страница 6: ...the footswitch to the appropriate connection point on the rear of the console noting the alignment of the orientation key see page 5 The footswitch can be disconnected for storage and to facilitate c...

Страница 7: ...e with ISO R 32 and ANSI NFPA 99 USA Ensure there is enough gas in the Cylinder prior to starting the procedure The console Cylinder Symbol indicates empty when the gas supply pressure drops below 350...

Страница 8: ...tion protocol before using a Cryo Tip probe Allow Cryo Tip probe to cool to room temperature after a sterilization procedure Before connecting the Cryo Tip probe inspect it for signs of obvious damage...

Страница 9: ...ure for the probe The pulsing may vary or cease completely depending on the pressure of gas in the cylinder If the probe performance drops below 100 permanently check the cylinder gas pressure or susp...

Страница 10: ...ns The Cryomatic system has the ability to detect a range of system faults In the unlikely event of a fault condition arising the fault symbol icon will flash and a short error message will be display...

Страница 11: ...rd Range Part No Description 2509 P 8000 2 5mm Standard Retinal Probe 2509 P 8001 2 5mm Extended Retinal Probe Special Range 2509 P 8002 2 5mm Mid Reach Retinal Probe 2509 P 8005 3mm Glaucoma 2509 P 8...

Страница 12: ...e Cryo Tip probe Instructions Point Of Use No particular requirements although excess soiling can be removed with disposable cloth paper wipe Containment Transportation Care should be taken to ensure...

Страница 13: ...ed When sterilising multiple probes in one sterilization cycle ensure that the manufacturer s maximum load is not exceeded Table 1 sterilizer type temperature pressure exposure time drying time pre va...

Страница 14: ...abrasive metal polish should be used b Cleaning the console surface c Inspecting the Cryo Tip probes for signs of damage before every use d Inspecting the footswitch and cord for signs of damage befor...

Страница 15: ...ptor N2 O Size VF 2508 P 7016 Cylinder Adaptor CO2 Size VF 2509 P 8009 Cylinder Adaptor N2 O US CGA326 2509 P 8012 Instructions For Use 2509 P 8011 High Pressure Hose 1m 2509 P 8014 High Pressure Hose...

Страница 16: ...contact supplier for repair Probe freezes but performance is poor Possible partial blockage in the probe possibly caused by excess moisture after sterilisation procedure Probe may appear to begin to...

Страница 17: ...in this warranty are expressly conditional upon the following i No alterations or repairs of any malfunction of the system shall be made to the system except by the manufacturer or his authorized repr...

Страница 18: ...Weight 2 5kg 6lbs Environmental Conditions Storage Operating Temperature Range 20 C to 50 C 10 C to 40 C Relative Humidity 10 to 80 30 to 70 Atmospheric Pressure 500hPa to 1060hPa 700hPa to 1060hPa Cl...

Страница 19: ...symbol contains an activity bar whenever the gas supply is being checked It is solid when reporting cylinder pressure Ready Symbol Displayed whenever the freeze function can be activated via the foot...

Страница 20: ...c System uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B...

Страница 21: ...rcial or hospital environment Surge IEC 61000 4 5 1kV differential mode 2kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupti...

Страница 22: ...ed by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with this symbol NOTE 1 At 80MHz and...

Страница 23: ...P 800MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at maximum output power not listed above the recommended separation...

Страница 24: ...of continued product improvement we reserve the right to alter specifications at any time without prior notice Keeler Instruments Inc 456 Parkway Broomall PA 19008 USA Toll Free 1 800 523 5620 Tel 610...

Отзывы: