7. Cleaning & Sterilisation
Instructions continued
Drying
No particular requirements.
Maintenance
Check for obvious signs of damage – return for manufacturer if any damage is noted.
Inspection & Function Testing
Visually inspect for damage and wear. Check probe tips for signs of bending, distortion or other damage.
Connect Cryo-Tip Probe to ‘Cryomatic’ console to check correct and smooth function of probe quick release coupling.
Packaging
No particular requirements.
Sterilisation
Disinfection is only acceptable as a precursor to full sterilisation for reusable surgical instruments. See Table 1 for
recommended sterilization parameters using equipment meeting relevant standards2. Sterilizer manufacturer
recommendations should always be followed. When sterilising multiple probes in one sterilization cycle, ensure
that the manufacturer’s maximum load is not exceeded.
Table 1
sterilizer type
temperature
pressure
exposure time
drying time
pre-vacuum
(porous load)
134-137°C
(
273-279
°F)
3 bar
28.5 PSI
3 minutes
30 minutes
gravity
121-124°C
(
250-255
°F)
-
15 minutes
-
2 HTM2010 or equivalent
Storage
Sterile, packaged Cryo-Tip probes should be stored in a designated, area that is well ventilated and provides protection
from dust, moisture and temperature/humidity extremes. Cryo-Tip probes should be coiled loosely during storage.
Cleaning the Console
The Cryomatic console can be cleaned using a disposable cloth soaked
in mild detergent and warm water. Do not use abrasive compounds
or pads. The user should avoid getting electrical parts wet during the
cleaning process.
Warning: Switch off the electrical supply to the console and disconnect
the power cord from the mains supply before cleaning and inspection.