Audio Pause
Reset
Blockage Alarm
Therapy Interrupted
WARNING:
Unit may be running, but therapy
is interrupted. Press ‘?’ for more information.
Blockage
Potential Blockage detected
Blockage Alert
Blockage
Continuous
0 mmHg
Exit
Press ‘?’ for more information
1. Ensure both clamps on the dressing and canister tubing are open.
2. Ensure that the tubing is not kinked, crimped or blocked in any way.
3. If the Blockage Alert remains after completing steps 1 and 2, lowering the therapy unit and tubing to the
same level as, or below, the wound site may resolve this alert. If the alert is cleared by lowering the
unit, normal use may resume.
4. Press ‘
’ on this screen to return to the ‘
.
To resolve this alarm:
Reset
Home Screen’
This alarm screen appears when the ActiV.A.C. Therapy Unit has
determined that a blockage is present.
This alarm will be accompanied by a repeating audible tone.
This alarm may be silenced for two (2) minutes during
troubleshooting by pressing
.
®
‘Audio Pause’
WARNING: Under clinician supervision, replace V.A.C. Dressing with alternate
dressing if therapy is interrupted for more than two hours.
®
Blockage Alarm Therapy Interrupted
Blockage Alert
1. Ensure both clamps on the dressing and canister tubing are open.
2. Ensure that the tubing is not kinked, crimped or blocked in any way.
3. If the Blockage Alert remains after completing steps 1 and 2, lowering the therapy unit and tubing to the
same level as, or below, the wound site may resolve this alert. If the alert is cleared by lowering the
unit, normal use may resume.
4. Press
to return to the
screen.
To resolve this alert:
‘Exit’
‘Home”
This al
screen appears when the
ert
ActiV.A.C. Therapy Unit has
detected a potential blockage and is working to determine if a
Blockage Alarm needs to be displayed.
This alert will be accompanied by a single audible tone.
®
Alerts and Alarms (cont.)
The ActiV.A.C. Therapy Unit continues to attempt to apply therapy.
®
If alarm condition cannot be resolved, contact KCI. See back cover of this User Manual
for country specific contact information.
WARNING: Therapy unit remains on; however, negative pressure at the wound
may be below set pressure, potentially compromising therapeutic benefits.