26
Intensity Control
·
·
·
·
Intensity is related to the time it takes to reach the target therapy level after the initiation of
therapy.
The lower the intensity setting, the slower the target therapy level will be reached.
It is recommended that new patients begin therapy at the lowest intensity setting as this
allows for slower increase of negative pressure once the foam is compressed in the
wound.
The intensity can remain at the minimum setting throughout the entire length of treatment, if
desired.
Intensity
Intensity
Intensity
Low
Medium
High
Press to change levels. Green crescent changes with each setting.
Default setting is: Low
From the Clinician Mode Home screen, press ‘
’,
then
to access
th
Therapy
then ‘
’,
‘
’
e Intensity button.
Next
Settings
Settings changed manually take immediate effect when therapy is on.
Settings (cont.)
Pressure Settings
Exit
Pressure
125 mmHg
+
_
Press ‘
’ to return to the ‘Settings’ screen.
Exit
From the Clinician Mode Home screen, press ‘
’,
then
to access th
Therapy
then ‘
’,
‘
’, then
e Pressure screen.
Next
Settings
‘Pressure’
Settings changed manually take immediate effect when therapy is on.
Use the
buttons to change to the desired pressure.
Pressure can be set from 25 to 200 mmHg in increments of 25 mmHg.
+
_
D
setting is: 125 mmHg
efault
26
®
A
ctI